CHAPTER 2 DECADE II DCC upgrade kit install guide, edition 5
31
3. Initialization
After upload of the new dual sensor FW the DECADE II should be reset to a
defined state by executing the following steps:
a. Clear existing time files:
A power-up of the detector while depressing
the F5 button will clear the existing time files in SRAM memory (so
hold F5 button and switch on detector)
b. Reset to factory settings:
this action will reset all non-volatile
parameters to their default values. To execute a factory reset go to
the CONFIG screen by pressing F1 in the MAIN menu:
Subsequently hold the <ENTER> button for 4 seconds. After a factory
reset please re-adjust the LCD contrast in the CONFIG screen to an
appropriate level (~9).
The DECADE II will malfunction if the reset procedures described in
point (a) and (b) are not executed. (embedded software can freeze or
detector remains in a reboot loop after power-up )
4. Check detection of the TC-board board by the DECADE II firmware
The LM 35 board is recognized by the software during start up of the DECADE
II. To check if the installation was successful start up the detector and enter the
DIAG mode from the MAIN menu. Subsequently, enter the SYSTEM menu by
pushing F5. In the SYSTEM menu, "Tsensor = LM + SMT", should be displ
ayed.
B
o
o
t
=
1
.
3 0
S
Y
S
T
E M
68
F
i
r
m
w
a
r
e
=
3
.
0 7
C
h
e
c
k
s
u m
=
4 2 3 1 2 4 1 5
T s e n s o
r
=
L
M
+
S
M
T
P R E V
If "Tsensor = SMT" is displayed, please restart the detector again.
If the TC-board is not correctly recognized, please re-check if the TC-
board and LM35 sensor cable are firmly fixed in their sockets.
T e m p
=
3 0 ° C
C o n t
r a s
t
=
2 0
C O N F I G
35
V a
l
v e =
p r e s e n t
V o u t
=
1 V F S
I D 1
m a s
t e r =
n o
P R E V
Содержание DECADE II
Страница 2: ......
Страница 8: ...6 ...
Страница 20: ...18 Installation guide Figure 7 Removal of sensor board from the connector panel ...
Страница 36: ...34 Appendix A Installation of 2nd internal cell cable ...
Страница 42: ......