24 Installation guide
Only a few DECADE II detectors on the market do not have this additional
LM35 sensor preinstalled (first series s/n, < 171.00025).
If your DECADE II does not have a pre-installed LM35 sensor, do not
procede with the installation and contact the manufacturer. The
DECADE II should be upgraded at the factory.
Fix the LM 35 connector on the socket, designated CON1, on the TC-board
as shown in figure 15.
Figure 16 TC board with mounted TC-board fixing connector.
Place the TC-board fixing connector (p/n 171.0690) on top of connector 4 on
the TC-board as shown in figure 16.
Use gentle force when installing the TC-board fixing connector (p/n
171.0690) on the TC-board. Prevent bending of the TC-board by holding
in place with your hands. during installation.
This connector fixes the TC-board between control and sensor board to
ensure that the board will not detach of the DECADE II control board.
Содержание DECADE II
Страница 2: ......
Страница 8: ...6 ...
Страница 20: ...18 Installation guide Figure 7 Removal of sensor board from the connector panel ...
Страница 36: ...34 Appendix A Installation of 2nd internal cell cable ...
Страница 42: ......