30 Installation guide
15. Upload of firmware version 3.xx
Upgrade the firmware of the DECADE II with an FW version which supports
dual sensor board control (FW > 3.20) as described in the FW upgrade user
guide (171.7032). A CD-ROM is provided with DECADE II Dialogue and
DECADE II firmware release 3.20 >. Please check for new releases on the
internet: http://www.antecleyden.com/support/. If a newer version of the FW
and Dialogue are available please install these versions.
If something goes wrong during the installation or upload procedure the
detector will start up with one of the error messages listed in chapter 4 “Error
messages” of the service manual.
In case of an error message, follow the corresponding instruction in the
table with error messages in the service manual.
16. Check & calibration of DECADE II DCC
1. Check FW version
After upload of the dual sensor FW restart detector and check if the FW
version displayed in the MAIN screen corresponds with the version number of
the uploaded anl file.
2. Check configuration of sensor boards
The presence of both boards can be simply checked in the DECADE II MAIN
screen.
A n
t
e c
L e y d e n
M A I N
1
2
D E C A D E -
I
I
L C - E C
w o
r
k s
t
a
t
i
o n
3
.
x x
C O N F
I G
D C
P U L S E
S C A N
D I A G
Toggle with the “+” / “-“ buttons between sensor board 1 and 2. The inverted
number in the top right corner should change from “1” to “2” corresponding to
sensor board 1 and 2. If the board number does not change it means that the
second sensor board is not properly recognized.
If the second sensor board is not recognized, re-check the dip switch
settings on both sensor boards and check if the sensor cable I
2
C/power
cable is connected properly.
Содержание DECADE II
Страница 2: ......
Страница 8: ...6 ...
Страница 20: ...18 Installation guide Figure 7 Removal of sensor board from the connector panel ...
Страница 36: ...34 Appendix A Installation of 2nd internal cell cable ...
Страница 42: ......