Performance Verification 3-4 ME7838A/E System Verification Procedure—Broadband System Configuration
VectorStar ME7838 Series MM
PN: 10410-00306 Rev. F
3-3
3-4
ME7838A/E System Verification Procedure—Broadband System
Configuration
The broadband system verification procedure is described below. The procedure assumes that the Broadband
VNA System Performance Verification Software has been installed to an External Personal Computer with
National Instruments GPIB interface running Microsoft Windows Operating System.
Equipment Required
•
Personal Computer:
•
Microsoft Windows Operating System
•
National Instruments GPIB interface
•
GPIB interface cable
•
Anritsu 3656B W1 Connector Calibration / Verification Kit
Special Precautions
When performing the procedures in this chapter, observe the following precautions:
•
Minimize vibration and movement of the system and attached components.
•
Clean and check pin depth and condition of all adapters and calibration components.
Procedure
1.
Using the GPIB interface cable to connect the external computer to the MS4647A/B rear panel system
GPIB connector. It is the upper GPIB port labeled
IEEE488.2 GPIB
.
2.
Dismount both 3743A Millimeter-Wave modules from the probe station, if required.
3.
Place both modules on a leveled surface so that both test ports are facing each other.
4.
Install the W1 female to female adapter from the 3656B Calibration / Verification Kit on the Port 1
3743A/E Millimeter-Wave Module. This converts Port 1 from a male test port to a female test port.
5.
Install the W1 male to female adapter from the 3656B Calibration / Verification Kit on the Port 2
3743A/E Millimeter-Wave Module.
6.
Run the Anritsu W1 Connector Broadband VNA System Performance Verification Software on the PC.
7.
Follow the directions displayed on the computer to perform calibration.
8.
Follow the directions displayed on the computer to perform measurement of the verification standards.
Note
The
3656B W1 Calibration / Verification Kit and System Performance Verification Software
User Guide – 10410-00286
explains in detail the Personal Computer requirements and procedures
to be used for the installation and operation of the verification software on the Personal Computer.
Note
Do not connect to the lower GPIB port labeled
Dedicated GPIB
.
Note
If the heatsink has been separated from the module while installing to the probe station, install the
module onto its original heatsink.
Содержание VectorStar ME7838 Series
Страница 2: ......
Страница 14: ...1 6 ME7838 Series VNA System Overview General Information 1 6 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 28: ...1 7 Recommended Test Equipment General Information 1 20 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 76: ...4 10 ME7838G System Operation Theory Of Operation 4 16 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 94: ...5 3 ALC Level Calibration Adjustment 5 18 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 176: ...A 3 ME7838G WR05 S Parameter Measurements Test Record Test Records A 4 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 177: ......