Theory Of Operation
4-8 ME7838E System Operation—Waveguide Band Configuration
VectorStar ME7838 Series MM
PN: 10410-00306 Rev. F
4-13
4-8
ME7838E System Operation—Waveguide Band Configuration
This section describes the system operation of the Waveguide Band ME7838E VNA System.
Stimulus Signal Generation
The MS464xB VNA outputs an RF signal to the 3739B or 3739C Test Set.
In the test set, the RF signal is then routed to the Port 1 RF Output or Port 2 RF Output via a transfer switch
after amplified and fed to the 3744E Millimeter-Wave Modules.
In the 3744E Millimeter-Wave Module, the frequency of the RF signal is either doubled or tripled and then
output to the DUT.
For operation between 56 GHz and 80 GHz, the frequency of the RF signal is doubled in the module. For
operation between 80 and 110 GHz, the frequency of the RF signal is tripled.
Test Signal Processing
The MS464xB VNA outputs two LO signals to the 3739B or 3739C Test Set. The LO signals are amplified (and
attenuated, if required). It is then routed to Port 1 LO Output or Port 2 LO Output and fed to the 3744E
Millimeter-Wave Modules.
In the 3744E Millimeter-Wave Module, a sample of the stimulus signal is down-converted to intermediate
frequency (IF) using the LO signal from the 3739B or 3739C Test Set. The output is the Reference IF signal.
The reflected/transmitted test signal received at the Waveguide test port of the 3744A module is coupled to the
test channel down-converter which converts the test signal to IF signal using the LO signal from the 3739B or
3739C Test Set. The output is the Test IF signal.
Both the Reference and Test IF signals are then fed to the 3739B or 3739C Test Set. The IF signals pass
through the 3739B or 3739C Test Set and are fed to the IF Inputs of the MS464xB VNA for further signal
processing.
Содержание VectorStar ME7838 Series
Страница 2: ......
Страница 14: ...1 6 ME7838 Series VNA System Overview General Information 1 6 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 28: ...1 7 Recommended Test Equipment General Information 1 20 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 76: ...4 10 ME7838G System Operation Theory Of Operation 4 16 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 94: ...5 3 ALC Level Calibration Adjustment 5 18 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 176: ...A 3 ME7838G WR05 S Parameter Measurements Test Record Test Records A 4 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 177: ......