5-3 ALC Level Calibration
Adjustment
5-16
PN: 10410-00306 Rev. F
VectorStar ME7838 Series MM
Port 1 RF Leveling Calibration
11.
Select
Back
and then
RF Leveling
. The
ALC [3739-RF]
menu as shown in
will
appear.
12.
Connect the Keysight W8486A power sensor to the Test Port of the Millimeter-Wave Module at Port 1.
13.
Select
Port 1 ALC Cal
.
14.
Click the
Start Cal
button to start the calibration.
15.
When the calibration is complete, disconnect the Keysight W8486A power sensor.
16.
Connect the Anritsu SC7770 power sensor to the Test Port of the 3743A/3744A Module at Port 1.
17.
Change
Freq Range
to
Bnd1:54G-70GHz
.
18.
Select
Port 1 ALC Cal
.
19.
Click the
Start Cal
button to start the calibration.
20.
After calibration is complete, disconnect the Anritsu SC7770 power sensor from the Millimeter-Wave
Module at Port 1.
Port 2 RF Leveling Calibration
21.
Connect the Anritsu SC7770 power sensor to the W1 Test Port of the Millimeter-Wave Module at Port 2.
22.
Select
Port 2 ALC Cal
.
23.
Click the
Start Cal
button to start the calibration.
Figure 5-12.
ALC [3739-RF] Menu
Содержание VectorStar ME7838 Series
Страница 2: ......
Страница 14: ...1 6 ME7838 Series VNA System Overview General Information 1 6 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 28: ...1 7 Recommended Test Equipment General Information 1 20 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 76: ...4 10 ME7838G System Operation Theory Of Operation 4 16 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 94: ...5 3 ALC Level Calibration Adjustment 5 18 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 176: ...A 3 ME7838G WR05 S Parameter Measurements Test Record Test Records A 4 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 177: ......