Troubleshooting
6-7 General Troubleshooting of the 3739A Test Set
VectorStar ME7838 Series MM
PN: 10410-00306 Rev. F
6-7
6-7
General Troubleshooting of the 3739A Test Set
This section provides general troubleshooting procedures of the 3739A Test Set.
Suggested Troubleshooting Steps
The suggested troubleshooting steps for 3739A Test Set are as follows:
•
Verify that the DC voltages from the power supply are at the expected level. Refer to
Test – 3739A Power Supply DC Check” on page 6-8
•
Verify that the DC bias voltages of RF components and fan assembly are present on the Test Set A1 Bias
Control PCB Assembly. Refer to
“Troubleshooting Test – 3739A A1 Bias Control PCB DC Bias Check”
.
•
Verify that the power level of the Port 1 RF port and Port 2 RF port are at the expected level. Refer to
“Troubleshooting – 3739A Test Set Port Power Level Check” on page 6-11
•
Verify that the power level of the Port 1 LO port and Port 2 LO port are at the expected level. Refer to
“Troubleshooting – 3739A Test Set Port Power Level Check” on page 6-11
Note
If RF port and LO port power levels are at the expected level, then the fault most likely lies in the
Millimeter-Wave Module that is connected to the respective port.
Содержание VectorStar ME7838 Series
Страница 2: ......
Страница 14: ...1 6 ME7838 Series VNA System Overview General Information 1 6 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 28: ...1 7 Recommended Test Equipment General Information 1 20 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 76: ...4 10 ME7838G System Operation Theory Of Operation 4 16 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 94: ...5 3 ALC Level Calibration Adjustment 5 18 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 176: ...A 3 ME7838G WR05 S Parameter Measurements Test Record Test Records A 4 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 177: ......