5-3 ALC Level Calibration
Adjustment
5-14
PN: 10410-00306 Rev. F
VectorStar ME7838 Series MM
ALC Calibration Procedure—Banded Millimeter-Wave Configuration
Port 1 IF Leveling Calibration
1.
Select
IF Leveling
. The
ALC [3739-IF]
menu as shown in
will appear.
2.
Connect the Anritsu SC7770 power sensor to the Test Port of the Millimeter-Wave Module at Port 1.
3.
Ensure that
Freq Range
displays
Bnd1:54G-70GHz
. If not, select
Freq Range
and select
Band 1
.
4.
Select
Port 1 ALC Cal
.
5.
Click the
Start Cal
button to start the calibration.
6.
After calibration is complete, disconnect the Anritsu SC7770 power sensor from the Millimeter-Wave
Module at Port 1.
7.
Change the
Freq Range
to
Bnd2:70G-125GHz
.
8.
Connect the Keysight W8486A power sensor to the Test Port of the Millimeter-Wave Module at Port 1.
9.
Select
Port 1 ALC Cal
.
10.
Click the
Start Cal
button to start the calibration.
Note
Remove the Waveguide Adapter prior to performing the ALC Calibration. Refer to the adapter
installation instructions in Chapter 5, Waveguide Adapter Kit Instructions, of
VectorStar
Broadband/Banded Millimeter-Wave Modules Reference Manual
, part number 10410-00311.
Содержание VectorStar ME7838 Series
Страница 2: ......
Страница 14: ...1 6 ME7838 Series VNA System Overview General Information 1 6 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 28: ...1 7 Recommended Test Equipment General Information 1 20 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 76: ...4 10 ME7838G System Operation Theory Of Operation 4 16 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 94: ...5 3 ALC Level Calibration Adjustment 5 18 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 176: ...A 3 ME7838G WR05 S Parameter Measurements Test Record Test Records A 4 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 177: ......