VectorStar ME7838 Series MM
PN: 10410-00306 Rev. F
8-1
Chapter 8 — Removal and Replacement
Procedures For 3739B and 3739C
8-1
Introduction
This chapter provides procedures for removing and re-installing the replaceable components and
sub-assemblies in the 3739B and 3739C Test Sets.
8-2
Disassembly Procedure – Power, Disconnect, and Covers
Use this procedure to prepare the 3739B or 3739C Test Set for removal and replacement procedures for all of
its replaceable components. Other than the front and rear panel cables, all replacement components require
this common disassembly procedure.
1.
Prepare a clean and static free work area. Make sure the work area is well grounded. Cover the work
surface with a soft, clean anti-static mat.
2.
Provide all personnel with appropriate anti-static grounding wrist straps and similar equipment.
3.
Power down the VNA.
4.
Power down the Test Set.
5.
Disconnect the rear panel cables between VNA and the Test Set. Disconnect the Power Cable from the AC
Mains.
6.
When the millimeter-wave modules were originally shipped, each module was calibrated and labeled
with the appropriate VNA Port Number. Make a note as to which port each module is connected.
7.
Disconnect the front panel cables between the VNA/Test Set and the Millimeter-Wave Modules.
8.
Set the Millimeter-Wave Modules and the Port-to-Module Note aside in a secure, clean, and anti-static
environment.
9.
Make sure all VNA front rear panel cables have been disconnected. Remove VNA from the top of the Test
Set.
10.
Set the VNA aside in a secure, clean, and anti-static environment.
11.
Move the Test Set to the repair area.
Caution
The weight of a fully equipped MS464xx VNA is greater than 28 kg (57.3 pounds). Use two (2)
or more people to lift and move the VNA. There is a risk of back injury if this instrument is
moved by one person. Make sure that any equipment carts can safely carry the instrument
weight.
Содержание VectorStar ME7838 Series
Страница 2: ......
Страница 14: ...1 6 ME7838 Series VNA System Overview General Information 1 6 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 28: ...1 7 Recommended Test Equipment General Information 1 20 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 76: ...4 10 ME7838G System Operation Theory Of Operation 4 16 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 94: ...5 3 ALC Level Calibration Adjustment 5 18 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 176: ...A 3 ME7838G WR05 S Parameter Measurements Test Record Test Records A 4 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 177: ......