Adjustment
5-3 ALC Level Calibration
VectorStar ME7838 Series MM
PN: 10410-00306 Rev. F
5-11
ALC Calibration Procedure—Broadband Configuration
1.
Select
IF/RF/Base ALC Cal
; the
ALC[3739-All]
.
2.
If the Millimeter-Wave Modules are MA25300A, change Module to
145GHz Capable
. If the
Millimeter-Wave Modules are MA25400A, change Module to
220GHz Capable
.
Port 1 ALC Calibration
3.
Confirm that Freq Range displays
Bnd1:54G-70GHz
. If not, select
Freq Range
and select
Band 1.
4.
Connect the Anritsu SC7770 power sensor to the Test Port of the Millimeter-Wave Module at Port 1.
5.
Select
Port 1 ALC Cal
.
6.
Click the
Start Cal
button to start the calibration.
7.
When calibration is complete, disconnect the Anritsu SC7770 power sensor and connect the Keysight
W8486A power sensor to the Test Port of the Millimeter Module at Port 1. If the Millimeter-Wave
Modules are MA25300A or MA25400A, remove the 33W.8F50 Adapter from the Anritsu SC7770 power
sensor and install the adapter to the 35WR10WF adapter mounted on the Keysight W8486A power
sensor before connecting it to the Test Port of the Millimeter Module at Port 1.
8.
Change the
Freq Range
to
Bnd2:70G-125GHz.
9.
Select
Port 1 ALC Cal
.
10.
Click the
Start Cal
button to start the calibration.
Figure 5-10.
ALC[3739-All] Menu
Содержание VectorStar ME7838 Series
Страница 2: ......
Страница 14: ...1 6 ME7838 Series VNA System Overview General Information 1 6 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 28: ...1 7 Recommended Test Equipment General Information 1 20 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 76: ...4 10 ME7838G System Operation Theory Of Operation 4 16 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 94: ...5 3 ALC Level Calibration Adjustment 5 18 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 176: ...A 3 ME7838G WR05 S Parameter Measurements Test Record Test Records A 4 PN 10410 00306 Rev F VectorStar ME7838 Series MM ...
Страница 177: ......