4-8 Spurious Signals Tests
Performance Verification
4-10
PN: 10370-10386 Rev. B
MG362x1A MM
Harmonic Test Procedure
1.
Determine the first test frequency to test based on the test record, options and model number.
2.
Set up the MG362x1A as follows:
a.
Reset the instrument by pressing
SYSTEM
, then
Reset
. After reset, the CW menu is displayed.
b.
Press
FREQUENCY
to open the current frequency parameter for editing and set F1 to the first
appropriate test frequency (based on model and options) in the test record.
c.
Press
LEVEL
to open the current power level parameter for editing.
d.
Set L1 to +10 dBm or to the maximum specified power level, whichever is less.
Refer to the MG362x1A Technical Data Sheet, PN: 11410-00928 on the Anritsu Internet site.
3.
Set up the spectrum analyzer as follows:
a.
Press the
AMPLITUDE
key. Press
REF LVL
and enter the current power level setting (L1) of the
MG362x1A.
b.
Press the
FREQUENCY
key and enter the current frequency setting (F1) of the MG362x1A.
c.
Press the
BW
key and press
MAN.
Set to the first appropriate test frequency’s RBW/VBW value
(based on model and options) in the test record.
d.
Press the
SPAN
key and set to the first appropriate test frequency’s SPAN value (based on model
and options) in the test record.
4.
Press the
PEAK SEARCH
key, then select
MARKER DELTA
.
5.
Press the
FREQUENCY
key on the spectrum analyzer and enter next harmonic frequency listed in the
test records.
6.
Press the
PEAK SEARCH
key,
7.
Record the reading from the spectrum analyzer into the test record.
8.
Repeat Step 5 through Step 7 for each of the harmonic frequencies listed in the test record.
9.
Press
FREQUENCY
on the MG362x1A to open the current frequency parameter for editing and set F1 to
the next appropriate test frequency (based on model and options) in the test record.
10.
On the SPA, press the
FREQUENCY
key and enter the current frequency setting (F1) of the MG362x1A.
11.
Determine if the power level for the MG362x1A and spectrum analyzer changes based on the new
frequency or options, if so, adjust the MG362x1A L1 and spectrum analyzer amplitude as needed.
12.
Press the
SPAN
key and enter the value listed in the test record for the appropriate test frequency.
13.
Press the
BW
key and enter the value listed in the test record for the appropriate test frequency.
14.
Repeat Step 4 through Step 13 for each of the test frequencies listed in the test record.
Содержание Rubidium MG362 1A Series
Страница 12: ...Contents 10 PN 10370 10386 Rev B MG362x1A MM ...
Страница 16: ...1 9 ESD Requirements General Information 1 4 PN 10370 10386 Rev B MG362x1A MM ...
Страница 30: ...2 6 RF Deck Assemblies Functional Description 2 14 PN 10370 10386 Rev B MG362x1A MM ...
Страница 66: ...3 11 FM and Phase Modulation Calibration Calibration 3 36 PN 10370 10386 Rev B MG362x1A MM ...
Страница 164: ...4 26 Pulse Modulation Tests with the 86100C Oscilloscope Performance Verification 4 98 PN 10370 10386 Rev B MG362x1A MM ...
Страница 292: ...B 5 Self Test Messages Instrument Messages B 8 PN 10370 10386 Rev B MG362x1A MM ...
Страница 293: ......