Troubleshooting
5-7 Self Test Messages
MG362x1A MM
PN: 10370-10386 Rev. B
5-15
5-7
Self Test Messages
.
System Messages that Display Fail
Procedure
1.
Perform a manual pre-calibration. Refer to
“Initial Calibration” on page 3-11
.
2.
Run self-test.
•
If no error message is displayed, the problem is cleared.
•
If any errors are displayed, go to step 3.
3.
Replace the B2 PCB, perform a manual pre-calibration and run self-test.
•
If no error message is displayed, the problem is cleared.
•
If any error messages are displayed, contact your local Anritsu Service Center for assistance.
5-8
SG Messages
.
MOD Supply Voltage Errors
Applies to units installed with modulation hardware Option 27.
Procedure
1.
Perform a manual pre-calibration. Refer to
“Initial Calibration” on page 3-11
.
2.
Run self-test.
•
If no error message is displayed, the problem is cleared.
•
If any errors are displayed, go to step 3.
3.
Replace the B3 PCB, perform a manual pre-calibration and run self-test.
•
If no error message is displayed, the problem is cleared.
4.
Perform modulation calibrations and verify performance.
If any error messages are displayed, contact your local Anritsu Service Center for assistance.
ALC Temperature Error.
Procedure
1.
Perform a manual pre-calibration. Refer to
“Initial Calibration” on page 3-11
.
2.
Run self-test.
•
If no error message is displayed, the problem is cleared.
•
If any errors are displayed, go to step 3.
3.
Replace the B5 PCB, perform a manual pre-calibration and run self-test.
•
If no error message is displayed, the problem is cleared.
4.
Perform level calibrations and verify performance.
Содержание Rubidium MG362 1A Series
Страница 12: ...Contents 10 PN 10370 10386 Rev B MG362x1A MM ...
Страница 16: ...1 9 ESD Requirements General Information 1 4 PN 10370 10386 Rev B MG362x1A MM ...
Страница 30: ...2 6 RF Deck Assemblies Functional Description 2 14 PN 10370 10386 Rev B MG362x1A MM ...
Страница 66: ...3 11 FM and Phase Modulation Calibration Calibration 3 36 PN 10370 10386 Rev B MG362x1A MM ...
Страница 164: ...4 26 Pulse Modulation Tests with the 86100C Oscilloscope Performance Verification 4 98 PN 10370 10386 Rev B MG362x1A MM ...
Страница 292: ...B 5 Self Test Messages Instrument Messages B 8 PN 10370 10386 Rev B MG362x1A MM ...
Страница 293: ......