5-8 SG Messages
Troubleshooting
5-16
PN: 10370-10386 Rev. B
MG362x1A MM
YIG Temperature Error
Procedure
1.
Perform a manual pre-calibration. Refer to
“Initial Calibration” on page 3-11
2.
Run self-test.
•
If no error message is displayed, the problem is cleared.
•
If any errors are displayed, go to step 3.
3.
Replace the B6 PCB, perform a manual pre-calibration and run self-test.
•
If no error message is displayed, the problem is cleared.
4.
Perform frequency calibrations and verify performance.
SWF Errors (Switched Filter Errors)
Procedure
1.
Perform a manual pre-calibration. Refer to
“Initial Calibration” on page 3-11
2.
Run self-test.
•
If no error message is displayed, the problem is cleared.
•
If any errors are displayed, go to step 3.
3.
Replace the B6 PCB, perform a manual pre-calibration and run self-test.
•
If no error message is displayed, the problem is cleared.
4.
Perform Level Cal calibrations and verify performance.
REG Errors (Regulator Errors)
Procedure
1.
Perform a manual pre-calibration. Refer to
“Initial Calibration” on page 3-11
2.
Run self-test.
•
If no error message is displayed, the problem is cleared.
•
If any errors are displayed, go to step 3.
3.
Replace the B1 module and perform a manual pre-calibration and run self-test.
•
If no error message is displayed, the problem is cleared.
4.
Perform all calibrations and verify performance.
The OCXO module is Not Phase-locked to the External 10 MHz Reference.
Procedure
1.
Perform a manual pre-calibration. Refer to
“Initial Calibration” on page 3-11
2.
Run self-test.
•
If no error message is displayed, the problem is cleared.
•
If any errors are displayed, go to step 3.
1.
Using a coaxial cable with BNC connectors, connect the rear panel
10 MHz REF IN
connector to the rear
panel
10 MHz REF OUT
connector.
2.
Disconnect the SMA cable.
Содержание Rubidium MG362 1A Series
Страница 12: ...Contents 10 PN 10370 10386 Rev B MG362x1A MM ...
Страница 16: ...1 9 ESD Requirements General Information 1 4 PN 10370 10386 Rev B MG362x1A MM ...
Страница 30: ...2 6 RF Deck Assemblies Functional Description 2 14 PN 10370 10386 Rev B MG362x1A MM ...
Страница 66: ...3 11 FM and Phase Modulation Calibration Calibration 3 36 PN 10370 10386 Rev B MG362x1A MM ...
Страница 164: ...4 26 Pulse Modulation Tests with the 86100C Oscilloscope Performance Verification 4 98 PN 10370 10386 Rev B MG362x1A MM ...
Страница 292: ...B 5 Self Test Messages Instrument Messages B 8 PN 10370 10386 Rev B MG362x1A MM ...
Страница 293: ......