Care, maintenance
Cleaning
InnuPure C12
Issue 01/2011
37
9
Care, maintenance
The InnuPure C12 is largely maintenance-free. Your care and maintenance
work is restricted to cleaning and changing fuses.
Warning!
You are not authorised to open the device hood.
The device hood may only be opened by the Technical Customer Service
department of Analytik Jena AG or by trained personnel!
Disconnect the mains plug before opening the device hood.
9.1 Cleaning
Danger! Short circuit!
Disconnect the mains plug from the InnuPure C12 connector prior to cleaning
the device with disinfectant.
The InnuPure C12 may only be operated again once it is completely dry.
•
Avoid soiling with careful handling of the sample substances.
•
Wipe away spilled samples or reagents immediately with an absorbent cloth
or paper towel.
•
If InnuPure C12 is used for the analysis of infectious material, special care
must be taken, as the InnuPure C12 cannot be decontaminated as a
complete unit.
•
Visible soiling must be removed immediately with the suitable means,
whereby it must be ensured that no solvent can enter the inside of the
device.
•
We recommend as suitable disinfectants:
Decosept AF spray disinfectant
Dr. Schuhmacher GmbH
Meliseptol HBV - cloths
B. Braun
Incidin Plus solution
•
If the InnuPure C12, with which infectious material has been processed, is
returned to Analytik Jena AG for servicing, decontamination is to be carried
out beforehand and is to be documented.
•
The magnets of the InnuPure C12 can be cleaned separately.
Insert the sample tray into the device.
Use the option 'Magnet out' to move the magnets out of the tip
adapters (
→ "
Remove the sample tray from device for easily cleaning (
→
sample tray", p. 25).
Clean the magnets using e.g. the above recommended disinfectants.
Содержание aj CYBERTRON InnuPure C12
Страница 1: ...InnuPure C12 Extraction system...
Страница 38: ...Up date of the InnuPure C12 Delete isolation protocols 36 Issue 01 2011 InnuPure C12...
Страница 42: ...Waste disposal 40 Issue 01 2011 InnuPure C12...
Страница 44: ...Technical data 42 Issue 01 2011 InnuPure C12...
Страница 46: ...Table of figures 44 Issue 01 2011 InnuPure C12...
Страница 47: ......