Directions for Use
– Bridle Pro
®
7
ENGLISH
PRO RANGE CLIP:
Ensure nasal tube fits securely. Place loose strand of bridle tubing
between the clear flats below the circular region of the clip.
PRO CLIP: 8, 10, 12FR
Ensure nasal tube fits securely in channel. Place loose strand of
bridle tubing in the clip’s hinge.
PRO CLIP: 14, 16, 18FR
Ensure nasal tube fits securely in channel. Place loose strand of
bridle tubing between the softer inner portion of the clip and the
outer, more rigid plastic section.
PRO RANGE CLIP:
Close the clip by gripping it near the base of the clip and firmly
snapping it closed.
PRO CLIP: 8, 10, 12FR
Close the clip by folding over the plastic edge and firmly
snapping it closed.
PRO CLIP: 14, 16, 18FR
Close the clip by folding over the plastic edge and firmly
snapping it closed.
After the clip has been fully closed, tie the two strands of bridle
tubing together in a simple knot below the clip.
Clip Opening Tool: If the clip must be opened, place the side of
the opening tool into the clip. Push in and turn slightly to open.
NOTE: Ensure proper position of the nasal tube, bridle tubing
and clip prior to closure.
ALSO NOTE: Some nasal tubes are more rigid than others and
may require a higher force to close clip.
WARNING: Do not use any foreign object to open or close the
clip as this may damage the clip, making it less secure and/or
ineffective. If damaged, use a new clip.
Cut the excess bridle tubing 1cm (1/2”) below the knot.
Note placement of the AMT Bridle Pro
®
in the patient’s chart.
WARNINGS: Throughout the useful life of the AMT Bridle Pro
®
,
the clip must be visually inspected for signs of damage or loose
attachment to the nasal tube. Monitor the clip and nasal tube
position regularly for nasal tube migration.
REMOVAL: To remove the AMT Bridle Pro
®
and nasal tube: cut
only ONE strand on the bridle tubing. Gently pull both the AMT
Bridle Pro and nasal tube out of the nose.
To remove only AMT Bridle Pro
®
: Cut only ONE strand of the
bridle tubing and open the clip. Gently pull on the opened clip
to remove the Bridle tubing from nose.
NOTE: After removal, dispose of the device following facility
protocol, local disposal guidelines, or through conventional
waste.
CAUTION: Great care must be taken to prevent accidental
swallowing as the bridle tubing can easily slide back into the
nose.
NOTE: Recommended for up to 30 days continuous use.
Diagrams Page 2-3:
19: Pro Range Clip
20: Nasal Tube
21: Clip
22: Pro Clip 8, 10, 12FR
23: Nasal Tube
24: Clip
25: Pro Clip 14, 16, 18FR
26: Nasal Tube
27: Clip
Diagrams Page 2-3:
28: 1 cm
29: Gently pull on bridle tubing ends to verify clip is closed
30: Hold tube fixed
Diagrams Page 2-3:
31: Cut excess bridle tubing
Diagrams Page 2-3:
32: Clip Opening Tool
Содержание Bridle Pro
Страница 3: ...AMT Headquarters Brecksville Ohio Made in America Cleveland Ohio ...
Страница 4: ...2 3 2 1 7 8 9 4 5 10 3 1 2 8 20 23 21 24 27 28 7 6 9 26 29 30 19 22 25 ...
Страница 5: ...3 6 5 4 10 11 12 16 17 18 11 12 13 14 15 31 32 ...
Страница 110: ...108 ...
Страница 111: ...www AppliedMedical net 109 APPLIED MEDICAL TECHNOLOGY INC ...