5-7
Operational Test
Uncrating/Installation Instructions
P-920012-041
NOTE: Access to monitor PROG or FUNC menus is protected by a
pass code. Only authorized persons or a qualified technician may
have access to them.
IMPORTANT
: Independent Monitor Access Code is 2030.
1. Insert a virgin disk into Independent Monitor.
IMPORTANT:
Use good quality floppy disks only.
2. Format the disk. Refer to
Operator Manual
(P-920012-042)
Section 5, "Floppy Disk," 'Formatting a Floppy Disk' for instruc-
tions to format a disk.
NOTE: Never remove a disk from the drive while the drive light is on.
3. Program a File name. Refer to Operator Manual (P-920012-042)
Section 5 "Floppy Disk," 'Programming a Filename.'
4. Set the Independent Monitor to Trigger Mode. Refer to
Operator
Manual
(P-920012-042) Section 5 "Floppy Disk," 'Starting Data
Recording.'
5. Press
CYCLE START
and allow cycle to run to completion.
6. To verify if independent monitor has recorded the information,
remove floppy disk. Refer to Section 5, "Floppy Disk," 'Starting
Data Recording.'
7. Remove floppy disk and verify if information was recorded.
8. Refer to
Operator Manual
(P-920012-042) for procedure to visu-
alize information on a PC.
9. Insert a virgin disk. Follow steps 1 to 3.
10. Call Department Supervisor or authorized person to communi-
cate passcode on independent monitor.
NOTE: We suggest to keep the same factory-set passcode.
IMPORTANT:
Keep your passcodes safe. If you lose them, there is
NO WAY to clear or reset them.
12. If unit is automated (Load/Unload Modules), refer to
Uncrating/
Installation Instructions Manual
(P-122994-142) to install con-
veyor modules.
This unit is now ready for a thorough System Field Test,
Calibration, and a first-time Preventive Maintenance Check-
list Inspection by a qualified STERIS service representa-
tive.
Contact your STERIS Service Representative for availabil-
ity and information on this service. Refer to the
Maintenance
Manual
(P-764330-313) for further details on these procedures.
Start-up Procedure for
Independent Monitor
Set recording to Trigger Mode
Program a Filename
Format the disk
Insert a virgin disk into
Independent Monitor
Independent Monitor Start-Up
Flowchart
Содержание RELIANCE 444 HTM
Страница 6: ...iv P 920012 041 Uncrating Installation Instructions This Page Intentionally Left Blank ...
Страница 8: ...vi P 920012 041 Uncrating Installation Instructions This Page Intentionally Left Blank ...
Страница 14: ...2 2 P 920012 041 Uncrating InstallationInstructions Uncrating Instructions Figure 2 1 Crate Tip Indicator ...
Страница 20: ...2 8 P 920012 041 Uncrating InstallationInstructions Uncrating Instructions This Page Intentionally Left Blank ...
Страница 36: ...5 8 P 920012 041 Uncrating Installation Instructions Operational Test This Page Intentionally Left Blank ...