3-2
P-920012-041
Uncrating/Installation Instructions
Installation Instructions
An Equipment Drawing, showing all utility and space requirements,
was sent to you after the order for this equipment was received. The
clearance space, specified on the Equipment Drawing is necessary
for proper installation, operation, and maintenance of the unit. If this
document is missing or misplaced, contact your STERIS represen-
tative, giving the serial, equipment, and model number of the unit.
A replacement copy will be sent promptly.
NOTE: If it is necessary to relocate controls on the unit, a special kit
is available. Contact STERIS technical support for further informa-
tion.
1. Clearance - Clearance space shown on Equipment Drawing is
necessary for easy installation and proper operation and main-
tenance of unit.
2. Barrier wall flange(s) installation - Refer to Equipment Drawing for
installation.
3. Utility service lines:
• To allow service on unit without shutting off building supply
lines, shutoff valves (not by STERIS) should be installed on
steam, air, and water lines to unit. Shutoff valves must be
capable of being locked in OFF position only.
• If unit is installed next to other equipment, shutoff valves and
disconnect switches should be located so that service can be
shut off to one piece of equipment at a time.
• The disconnect device of the equipment must be within easy
reach of the operator (preferably no more than 1 m away from
equipment).
• Utility service requirements are shown on Equipment Drawing.
4. Electricity
• A
disconnect switch meeting the requirements of EN-60947
(not by STERIS) should be installed in electric supply lines near
unit. Check Equipment Drawing for required amperage.
• Disconnect switches must be marked as the disconnecting
device for the equipment and must be capable of being locked
in
OFF
position only.
• If supply is 380 V or 400 V
, locate the 600 VA transformer,
inside electrical supply box, disconnect brown wire from 250 V
tap connection. Connect brown wire of primary side to 230 V
tap connection.
• Check Equipment Drawing or Nameplate (see Figure 2-2) for
proper voltage and amperage.
• Run HAR-approved supply cord (isolation 80°C minimum)
from disconnect switch to unit electrical supply box. A European-
approved type inlet bushing (strain relief) must be fitted to
opening on electrical supply box to protect supply cord.
5. A protective conductor terminal (ground) is required (Class 1
Equipment).
Installation
Requirements
Содержание RELIANCE 444 HTM
Страница 6: ...iv P 920012 041 Uncrating Installation Instructions This Page Intentionally Left Blank ...
Страница 8: ...vi P 920012 041 Uncrating Installation Instructions This Page Intentionally Left Blank ...
Страница 14: ...2 2 P 920012 041 Uncrating InstallationInstructions Uncrating Instructions Figure 2 1 Crate Tip Indicator ...
Страница 20: ...2 8 P 920012 041 Uncrating InstallationInstructions Uncrating Instructions This Page Intentionally Left Blank ...
Страница 36: ...5 8 P 920012 041 Uncrating Installation Instructions Operational Test This Page Intentionally Left Blank ...