5-3
Operational Test
Uncrating/Installation Instructions
P-920012-041
STATUS
STATUS
STATUS
STATUS
a. Remove lower service access panel on load and on unload
sides.
b. From load side, note the direction of the arrow on the motor
housing. Motor should turn in that direction.
c. On unload side, press the
CYCLE/START
, then, press
STOP/
RESET
touch pads to verify if motor shaft is rotating in
direction indicated by the arrow on the motor housing.
d. If pump is rotating in the wrong direction, turn Main Power
Switch to
OFF.
Position main disconnect switch to
OFF
.
e. On Main Electric Box, reverse L1 and L2 3-phase wires and
recheck shaft rotation.
f. Restore main disconnect switch to
ON
.
g. Turn Main Power Switch to
ON
position. Control will go through
initializing sequence.
h. Press
CYCLE START
. Repeat steps b and c. If pump rotates
in the right direction, allow cycle to run for 1 min. Press
STOP/
RESET
twice to abort cycle.
22. On load side, press
DOOR OPEN
touch pad. Display shows:
ACCESS CODE KNOWN?
NO
23. Use
VALUE
(up or down) touch pad to select: YES.
24. Press
CHANGE VALUES
touch pad. Display shows:
ENTER OPERATOR
ACCESS CODE: 0000
25. Press
CHANGE VALUES
touch pad.
26. On load side, press
EMERGENCY STOP BUTTON
. Interior
light and control turn
OFF
. Water is drained from washer.
27. Release
EMERGENCY STOP BUTTON
. Interior light is
ON
,
Control starts. Display shows:
MONITOR INITIALIZING
then..
AMSCO RELIANCE
MODEL 444
followed by the first cycle menu screen:
WARNING - FALL HAZARD: To pre-
vent falls, keep floors dry. Promptly
clean up any spills or drippage.
CAUTION: Once three-phase power
is connected, check pump for cor-
rect rotation (indicated by arrow on
pump motor). Incorrect pump rota-
tion may result in pump damage
and improper cleaning action.
CAUTION: Before operating unit, al-
ways position each manifold or ac-
cessory header over a manifold
connector. If manifolds or acces-
sory headers are not positioned
correctly, damage may result and
unit will be unable to effectively
wash loads.
Содержание RELIANCE 444 HTM
Страница 6: ...iv P 920012 041 Uncrating Installation Instructions This Page Intentionally Left Blank ...
Страница 8: ...vi P 920012 041 Uncrating Installation Instructions This Page Intentionally Left Blank ...
Страница 14: ...2 2 P 920012 041 Uncrating InstallationInstructions Uncrating Instructions Figure 2 1 Crate Tip Indicator ...
Страница 20: ...2 8 P 920012 041 Uncrating InstallationInstructions Uncrating Instructions This Page Intentionally Left Blank ...
Страница 36: ...5 8 P 920012 041 Uncrating Installation Instructions Operational Test This Page Intentionally Left Blank ...