2-1
Uncrating Instructions
Uncrating/Installation Instructions
P-920012-041
UNCRATING INSTRUCTIONS
2
NOTE: Unit weighs approximately 1000 lbs (454 kg). Use a forklift
to move the unit.
NOTE: Uncrate on level floor as close to installation site as possible.
NOTE: Operator Manual is enclosed in an envelope taped on the
side of the unit. Make sure appropriate person receives Operator
Manual.
1. Check Tip Indicator, located on upper left side of crate (see
Figure 2-1). Tip Indicator contains a blue compound at the
bottom of the indicator. If unit has been tipped, residue from the
blue compound will be found higher up in the indicator.
If unit has been tipped, contact your STERIS representative to
determine if a service technician is required to inspect the
equipment and determine if unit was damaged.
2. Position unit to open wooden crate from top and side. Provide a
clear work area on all sides.
3. Carefully cut and remove bands from crate. Discard bands
before continuing.
4. Remove and discard polythene wrap from crate.
5. Using a nail puller, remove and discard top wooden panel.
6. Using a nail puller, remove and discard side wooden panels.
7. Using a nail puller, remove and discard wooden crate frame.
8. Carefully remove barrier wall flanges and set aside.
9. Remove and discard shrink wrap and polystyrene wrap from
unit.
NOTE: Do not remove protective adhesive paper from the exterior
cabinet panels until after utilities are connected.
10. Open lower service access door and check tip indicator,
located inside lower service compartment (see Figure 2-2).
Tip'N Tell indicator contains a blue compound at the bottom of
the indicator. If unit has been tipped, residue from the blue
compound will be found higher up in the indicator.
If unit has been tipped, contact your STERIS representative to
determine if a service technician is required to inspect the
equipment and determine if unit was damaged.
11. Close doors to the lower service access and chemical storage
compartments.
NOTE: Do not open chamber door(s) until procedures in "Prepare
Equipment for Installation," later in this section, are performed.
Open Crate
WARNING - PERSONAL INJURY
AND/OR EQUIPMENT DAMAGE
HAZARD: When moving the unit,
use a forklift. Do not use door(s) or
door handle(s) for lifting.
WARNING - LACERATION AND/OR
EYE INJURY HAZARD: When re-
moving bands, wear gloves and
eye protection, and always use a
tool specifically designed to cut
the bands. The bands used to
secure this carton can cause
personal injury when cut and ten-
sion is released.
Содержание RELIANCE 444 HTM
Страница 6: ...iv P 920012 041 Uncrating Installation Instructions This Page Intentionally Left Blank ...
Страница 8: ...vi P 920012 041 Uncrating Installation Instructions This Page Intentionally Left Blank ...
Страница 14: ...2 2 P 920012 041 Uncrating InstallationInstructions Uncrating Instructions Figure 2 1 Crate Tip Indicator ...
Страница 20: ...2 8 P 920012 041 Uncrating InstallationInstructions Uncrating Instructions This Page Intentionally Left Blank ...
Страница 36: ...5 8 P 920012 041 Uncrating Installation Instructions Operational Test This Page Intentionally Left Blank ...