7-1
Cycle and Control Value Programming
Operator Manual
P129394-068
AMSCO
®
400 Series Medium Steam Sterilizers are shipped with
factory-set cycles, cycle values and control values programmed into
the control (see
, page ii). These are the cycles
and values to which the control will default should a memory failure
ever occur. These preset values can be changed to tailor the
sterilizer to the operating environment in which it has been placed.
To change these values, refer to the Change Values Procedures later
in this section.
NOTE:
If a memory failure occurs, operator-set cycles and values
are lost. Factory-set cycles, cycle values and control values appear
on the display when the sterilizer power is switched on.
Important:
Applicable cycles have been validated to satisfy the
(page ii). If cycle parameters
(sterilize time, dry time, temperature) other than those listed are
required, it is the responsibility of the healthcare facility to consult
and follow the device manufacturer’s written instructions.
NOTE:
Although AMSCO 400 Series Medium Steam Sterilizers are
factory-set to operate using Fahrenheit temperature units, the
sterilizer can be reprogrammed to display and print using Celsius
temperature units.
7.1 Change Values
The Change Values procedure described in the following pages can
be used to adjust cycle values within a limited range and to make
selections affecting the operating style of the unit. Refer to
for cycle values.
NOTE:
Default cycle values are shown on
The Change Values Procedure can be used to adjust the following
values:
•
Name: The user can select between 3 to 4 factory-programmed
names (depending on sterilizer configuration), or can create up
to two user-defined cycle names.
•
Sterilization Exposure Temperature: Exposure temperatures can
be adjusted, but within limits. Refer to
for details.
•
Sterilization Exposure Time (
Important:
This value is limited to a
minimum value, refer to
). Refer to
for range limits.
•
Dry Time. Refer to
for range limits.
NOTE:
Change Values can also be used to make adjustments to the
Change Machine Setup values shown on
The Change Values touch screens are accessed by pressing the
Change Values touchpad (see Screen #2). All Change Values
options can be secured (or locked out) using a supervisor’s access
code. It is recommended that supervisor’s Access Code be used to
prevent unauthorized personnel from changing cycle and system
setup parameters.
IMPORTANT
Applicable cycles have been validated
to satisfy the requirements outlined in
(page ii). If cycle parameters
(sterilize time, dry time, temperature)
other than those listed are required, it
is the responsibility of the healthcare
facility to consult and follow the device
manufacturer’s written instructions.
CYCLE AND CONTROL VALUE PROGRAMMING
7
Содержание 36 H
Страница 2: ......
Страница 14: ...xii P129394 068 Operator Manual Techniques of Sterilization LIST OF FIGURES Descrption Page ...
Страница 59: ...6 11 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 63: ...6 15 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 66: ...6 18 P129394 068 Operator Manual Sterilizer Operation Figure 6 9 Typical Printout of a Liquid Cycle ...
Страница 67: ...6 19 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 78: ...6 30 P129394 068 Operator Manual Sterilizer Operation Healthcare Capital Equipment ...
Страница 116: ...8 10 P129394 068 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 140: ...9 24 P129394 068 Operator Manual Troubleshooting Healthcare Capital Equipment ...
Страница 152: ...10 12 P129394 068 Operator Manual Service Procedures Healthcare Capital Equipment ...