![Amsco 36 H Скачать руководство пользователя страница 137](http://html1.mh-extra.com/html/amsco/36-h/36-h_operators-manual_2936188137.webp)
9-21
Troubleshooting
Operator Manual
P129394-068
Alarm
Description
Screen with Service Instructions
Door Switch Failure
(Continued)
STATUS . . SERVICE INFORMATION:
238
DOOR SWITCH FAILURE
-> DOOR SWITCH OPEN WHILE SEAL SWITCH CLOSED
CAUSES AND CORRECTIONS:
1. DOOR SWITCH MALFUNCTION
-> CHECK LS1 (LS2) CONNECTIONS
-> READJUST LS1 (LS2)
-> REPAIR LS1 (LS2)
2. SEAL PRESSURE SWITCH MALFUNCTION
-> READJUST PS1 (PS2)
-> REPAIR LPS1 (PS2)