![Amsco 36 H Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/amsco/36-h/36-h_operators-manual_2936188065.webp)
6-17
Sterilizer Operation
Operator Manual
P129394-068
6.11 Liquid Cycle
This cycle is used for sterilizing liquids in vented closures.
1.
Refer to
2.
See instructions for using the loading car/transfer carriage or
rack and shelves, earlier in this section, for procedures to load
the sterilizer.
3.
Press
LIQUID
touch-screen pad to start the Liquid cycle. Refer
to
for descriptions of cycle use.
NOTE:
If the wrong cycle has been selected, see
4.
Sterilizer automatically progresses through cycle, as follows:
NOTE:
The control display indicates cycle progress in two ways:
a.
Large numerals in the screen’s center estimate remaining
cycle time,
b.
A change of phase name in the top status line of the screen.
ACTIVATE SEAL – Steam enters the door seal, pressing it
against inside surface of the door.
PURGE – Chamber is purged with steam. Start of condition is
printed.Countdown timer on the display is estimated; timer self-
corrects estimated time at the beginning of each phase.
CHARGE – Chamber is charged with steam. Start of steam
charge is printed.
STERILIZE – Start of sterilize exposure is printed when the
chamber reaches sterilization temperature. Chamber
temperature is printed every five minutes.
SLOW EXHAUST – Start of exhaust is printed and chamber is
exhausted to 0 psig.
SLOW EVACUATE – A vacuum is slowly drawn in the chamber to
5.0 inHg. This phase assures that the chamber is cooled to
208°F/95°C.
VAPOR REMOVAL – Filtered air enters the chamber at the same
time a vacuum is drawn to cool the load. Chamber pressure is
maintained between 1 and 5 inHg. Phase continues for 10
minutes.
AIR BREAK – Chamber is returned to atmospheric pressure.
RETRACT SEAL – A vacuum is drawn on the seal, retracting it
from inner surface of door.
COMPLETE – Complete tone sounds. Cycle summary and end of
cycle messages are printed.
5.
Unload sterilizer (see instructions for using the loading car/
transfer carriage or rack and shelves, earlier in this section for
procedures to remove load).
WARNING – EXPLOSION
HAZARD: This sterilizer is not
designed to process
flammable compounds.
WARNING – PERSONAL
INJURY HAZARD: Avoid
personal injury from bursting
bottles. Liquid sterilization
cycle must only be used for
liquids in borosilicate (Pyrex)
flasks with vented closures.
WARNING: It is
inappropriate for a
healthcare facility to
sterilize liquids for direct
patient contact.
WARNING – BURN
HAZARD: When sterilizing
liquids, you must observe
the following procedures:
•
Use Liquid cycle only.
•
Use only vented closures.
•
Use only Type I
borosilicate glass bottles.
•
Do not allow hot bottle to
be jolted.
•
Steam may be released
from the chamber when
door is opened. Step back
from the sterilizer each
time the door is opened to
minimize contact with
steam vapor.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE
HAZARD: Sterilization of
chloride-containing
solutions (e.g., saline) can
cause chamber corrosion
and is not recommended
by the manufacturer. If,
however, chloride-
containing solutions must
be processed, clean the
chamber after each use.
!
!
!
!
!
Содержание 36 H
Страница 2: ......
Страница 14: ...xii P129394 068 Operator Manual Techniques of Sterilization LIST OF FIGURES Descrption Page ...
Страница 59: ...6 11 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 63: ...6 15 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 66: ...6 18 P129394 068 Operator Manual Sterilizer Operation Figure 6 9 Typical Printout of a Liquid Cycle ...
Страница 67: ...6 19 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 78: ...6 30 P129394 068 Operator Manual Sterilizer Operation Healthcare Capital Equipment ...
Страница 116: ...8 10 P129394 068 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 140: ...9 24 P129394 068 Operator Manual Troubleshooting Healthcare Capital Equipment ...
Страница 152: ...10 12 P129394 068 Operator Manual Service Procedures Healthcare Capital Equipment ...