![Amsco 36 H Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/amsco/36-h/36-h_operators-manual_2936188054.webp)
6-6
P129394-068
Operator Manual
Sterilizer Operation
11. Close chamber door.
12. Sterilizer is now ready to run a cycle. Proceed to cycle
descriptions found later in this section of the manual.
6.6 Loading Car
Instructions:
Unloading
1.
Open chamber door.
2.
Move transfer carriage forward until transfer carriage aligns with
guide assembly (at floor).
3.
Latches on transfer carriage engage with bracket on chamber
end ring.
4.
Verify that transfer carriage is latched to chamber end ring by
pulling transfer carriage backward (transfer carriage should
remain stationary).
5.
Once transfer carriage is securely latched, grasp the loading car
handle and carefully pull loading car from chamber onto transfer
carriage until transfer carriage latch engages to loading car.
6.
Disengage transfer carriage latches from track inside chamber
by pushing carriage latch knob.
7.
Close the chamber door.
8.
Transfer load from sterilizer area.
6.7 Loading/Unloading
Sterilizer:
Rack and Shelves
If sterilizer is equipped with the rack and shelves option, refer to
instructions below and
Figure
.
1.
Open chamber door.
2.
Transfer load to shelves in chamber. Shelves slide out halfway to
facilitate loading.
3.
After loading the shelves, slide them to closed position to verify
shelf does not interfere with door operation (both doors if double-
door sterilizer).
4.
Close chamber door(s).
5.
The sterilizer is now ready to run a cycle. Refer to appropriate
cycle description in
S
ECTION
6
of this manual.
WARNING – BURN HAZARD:
» Sterilizer, rack/shelves and
loading car will be hot after
cycle is run. Always wear
protective gloves and apron
when removing a processed
load. Protective gloves and
apron must be worn when
reloading sterilizer following
the previous operation.
» Steam may be released from
the chamber when door is
opened. Step back from the
sterilizer each time the door
is opened to minimize
contact with steam vapor.
WARNING – PERSONAL
INJURY and/or EQUIPMENT
DAMAGE HAZARD: When
closing the chamber door,
keep hands and arms out of
the door opening and make
sure opening is clear of
obstructions.
!
!
WARNING – BURN HAZARD:
Sterilizer, rack/shelves, and
loading car will be hot after
cycle is run. Always wear
protective gloves and apron
when removing a processed
load. Protective gloves and
apron must be worn when
reloading sterilizer following
the previous operation.
!
Содержание 36 H
Страница 2: ......
Страница 14: ...xii P129394 068 Operator Manual Techniques of Sterilization LIST OF FIGURES Descrption Page ...
Страница 59: ...6 11 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 63: ...6 15 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 66: ...6 18 P129394 068 Operator Manual Sterilizer Operation Figure 6 9 Typical Printout of a Liquid Cycle ...
Страница 67: ...6 19 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 78: ...6 30 P129394 068 Operator Manual Sterilizer Operation Healthcare Capital Equipment ...
Страница 116: ...8 10 P129394 068 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 140: ...9 24 P129394 068 Operator Manual Troubleshooting Healthcare Capital Equipment ...
Страница 152: ...10 12 P129394 068 Operator Manual Service Procedures Healthcare Capital Equipment ...