Table 6-3. Factory-Set Optional Liquid Cycle and Cycle Values
The AMSCO 400 Series Medium Steam Sterilizer is equipped with the following optional factory programmed liquid
sterilization cycle and cycle values.
Cycles
Sterilize Temp.
Sterilize
Time
Dry
Time
Recommended Load
(by Chamber Size)
Validation
Standard
LIQUID
*
250°F
(121°C)
45 Min.
N/A
ST-8: 2008
*
Liquids sterilized using the Liquid cycle are not for use with patients.
Figure 6-8. Liquid Cycle In-Cycle Screens
Cycle Selection Screen
Cycle Selection Screen
Condition Phase
Sterilize Phase
Exhaust Phase
NOTE:
Countdown time on display is estimated; timer self-corrects estimated time at beginning
of each phase.
Press for More Cycles
Содержание 36 H
Страница 2: ......
Страница 14: ...xii P129394 068 Operator Manual Techniques of Sterilization LIST OF FIGURES Descrption Page ...
Страница 59: ...6 11 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 63: ...6 15 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 66: ...6 18 P129394 068 Operator Manual Sterilizer Operation Figure 6 9 Typical Printout of a Liquid Cycle ...
Страница 67: ...6 19 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 78: ...6 30 P129394 068 Operator Manual Sterilizer Operation Healthcare Capital Equipment ...
Страница 116: ...8 10 P129394 068 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 140: ...9 24 P129394 068 Operator Manual Troubleshooting Healthcare Capital Equipment ...
Страница 152: ...10 12 P129394 068 Operator Manual Service Procedures Healthcare Capital Equipment ...