![Amsco 36 H Скачать руководство пользователя страница 82](http://html1.mh-extra.com/html/amsco/36-h/36-h_operators-manual_2936188082.webp)
7-4
P129394-068
Operator Manual
Cycle and Control Value Programming
Table 7-1. Adjustment Ranges for Cycle Values
Cycle Type
Sterilize Time
*
* Sterilize time limits:
»
Prevac or SFPP cycles set to a temperature of 270-274°F (132-134°C) are limited to a four-minute
sterilize time.
»
Prevac or SFPP cycles set to a temperature of 275°F (135°C) are limited of a three-minute minimum
sterilize time.
Sterilize Temperature
Dry Time
270°F (132°C) Prevac
4 - 99 minutes
270 - 274°F (132 - 134°C)
1 - 99 minutes
275°F (135°C) Prevac
3 - 99 minutes
275°F (135°C)
250°F (121°C) Gravity
30 - 99 minutes
250 or 269°F (121 - 131°C)
0 - 99 minutes
270°F (132°C) Gravity
15 - 99 minutes
270°F (132°C)
250°F (121°C) Liquid
45 - 99 minutes
250°F (121°C)
10-Minute Vapor
Removal
(no user adjustment)
Test Cycles
Dart (Bowie-Dick Test)
Dart Warm-Up
Leak Test
Values Cannot Be Adjusted
Содержание 36 H
Страница 2: ......
Страница 14: ...xii P129394 068 Operator Manual Techniques of Sterilization LIST OF FIGURES Descrption Page ...
Страница 59: ...6 11 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 63: ...6 15 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 66: ...6 18 P129394 068 Operator Manual Sterilizer Operation Figure 6 9 Typical Printout of a Liquid Cycle ...
Страница 67: ...6 19 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Страница 78: ...6 30 P129394 068 Operator Manual Sterilizer Operation Healthcare Capital Equipment ...
Страница 116: ...8 10 P129394 068 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 140: ...9 24 P129394 068 Operator Manual Troubleshooting Healthcare Capital Equipment ...
Страница 152: ...10 12 P129394 068 Operator Manual Service Procedures Healthcare Capital Equipment ...