M470039-01 REV-F
Page 141
3)
Pressure Requirements:
The Inlet Pressure or pressure differential should never exceed 70PSI. Please see below for additional
precautions.
A)
Install a Pressure Regulator on the inlet end of the Water Cooling System. This will help regulate the
inlet water pressure. The Honeywell DialSet® Pressure Regulating Valve is a good reference.
B)
There may be circumstances where the water valve is latched open due to excessive inlet pressure and
GPM Flow. To identify this issue, simply monitor the water valve control system and listen for water
flow once the valve indicator LED has darkened. The water flow should be impeded. If flow
continues, reduce GPM flow and Inlet Pressure and continue to run test application. Monitor the front
panel temperature reading (Go to Menu
→
Display
→
Meter Mode
→
P+T
) and ensure it is operating
below 40°C.
Содержание Amrel PLA 2U
Страница 2: ......
Страница 4: ...M470039 01 REV F Page 3 This is page intentionally left blank...
Страница 6: ...M470039 01 REV F Page 5 This page intentionally left blank...
Страница 8: ...M470039 01 REV F Page 7 This page intentionally left blank...
Страница 132: ...M470039 01 REV F Page 131 SEVEN SPECIFICATIONAND POWER CURVE COMMON SPECIFICATIONS...
Страница 141: ...M470039 01 REV F Page 140...
Страница 145: ...M470039 01 REV F Page 144 2 Unplug the 12VDC connector 4 pin remove the hexagonal nut and take off the coil assembly...
Страница 146: ...M470039 01 REV F Page 145 3 Remove the 4 screws securing the solenoid and take off the solenoid assembly...
Страница 152: ...M470039 01 REV F Page 151 6 OUTLET BOTTOM interconnect pipe x 1pcs 7 Plastic Round Shim ID OD 0 003 Thick x 4 pcs...
Страница 161: ......
Страница 162: ......