M470039-01 REV-F
Page 108
Use “
” and “
” until you see MAIN CR2 parameter. Press enter:
CR1
LO= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CR1
LO= CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER.
CR1
HI= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CR1
HI= CNT:
Input mea sured volta ge from Curr.DMM then press ENTER key. A “SUCCESS” messa ge
should be display on the f ront panel LCD.
CONSTANT POWER (CP) LOW/MIDDLE/HIGH RANGE, MAIN LEVEL CALIBRATION
CP ma in leve l ca libra tion requires a power supply tha t ha s enough power output. Fo r exa mple, a PL W6K-60-150 model
needs a t lea st 40V/80A power output to ca libra te the unit.
Use “
” and “
” until you see MAIN CP0 parameter. Then press enter:
CP0
LO= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CP0
LO= CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER.
CP0
HI= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CP0
HI= CNT:
Input mea sured volta ge from Curr.DMM then press ENTER key. A “SUCCESS” messa ge
should be display on the f ront panel LCD.
Use “
” and “
” until you see MAIN CP1 parameter. Press enter:
CP1
LO= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CP1
LO= CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER.
CP1
HI= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CP1
HI= CNT:
Input mea sured volta ge from Curr.DMM then press ENTER key. A “SUCCESS” messa ge
should be display on the f ront panel LCD.
Use “
” and “
” until you see MAIN CP2 parameter. Press enter:
CP2
LO= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CP2
LO= CNT:
Input mea sured volta ge from Curr.DMM, then press ENTER.
CP2
HI= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CP2
HI= CNT:
Input mea sured volta ge from Curr.DMM then press ENTER key. A “SUCCESS” messa ge
should be display on the f ront panel LCD.
CV TRANCIENT LEVEL CALIBRATION
CV tra nsient leve l ca libra tion requires a power supply tha t covers full ra nge of input volta ge of loa d a nd is ca pa ble of
operating in CV a nd CC mode. Estima te the power consumption of eLoa d, beca use power supply will go to CC mode
while ca libra ting the loa d. Adjust the CC current level to a sa fe limit for loa d.
Use “
” and “
” until you see TRAN CV0 parameter. Then press enter:
CV0
LO= CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER.
CV0 HI = CNT:
Input measured voltage from Volt.DMM then press ENTER key. A “SUCCESS”
should be displa y on the front pa nel LCD.
Press “
” to select TRAN CV1. Then press enter.
CV1
LO=
CNT:
Input mea sured volta ge from Volt.DMM, then press ENTER
CV1
HI = CNT:
Input measured voltage from Volt.DMM then press ENTER key. A “SUCCESS” message
Содержание Amrel PLA 2U
Страница 2: ......
Страница 4: ...M470039 01 REV F Page 3 This is page intentionally left blank...
Страница 6: ...M470039 01 REV F Page 5 This page intentionally left blank...
Страница 8: ...M470039 01 REV F Page 7 This page intentionally left blank...
Страница 132: ...M470039 01 REV F Page 131 SEVEN SPECIFICATIONAND POWER CURVE COMMON SPECIFICATIONS...
Страница 141: ...M470039 01 REV F Page 140...
Страница 145: ...M470039 01 REV F Page 144 2 Unplug the 12VDC connector 4 pin remove the hexagonal nut and take off the coil assembly...
Страница 146: ...M470039 01 REV F Page 145 3 Remove the 4 screws securing the solenoid and take off the solenoid assembly...
Страница 152: ...M470039 01 REV F Page 151 6 OUTLET BOTTOM interconnect pipe x 1pcs 7 Plastic Round Shim ID OD 0 003 Thick x 4 pcs...
Страница 161: ......
Страница 162: ......