8-6 | 3050 RM Moisture Analyzer
Moisture Generator
Removing the moisture generator
1. Remove the top cover from the analyzer.
2. Remove the cover for the oven assembly. Facing the front of the
3050-RM analyzer, the moisture generator is located left- center on the
sample manifold plate, above the pressure transducer, between the
flowmeter and the sensor assembly. See Figure 8-2.
3. Disconnect the 1/16 inch fittings on all four sample tubes, which are
located on the bracket in the center of the sample oven (Figure 8-5).
4. Remove the four nuts that secure the sample manifold plate (the
square aluminum mounting plate, located in the front section of the
oven), and gently lift the manifold out of the oven (Figure 8-5).
All of the wiring connections should still be intact at this time; you
will need to rest the sample manifold plate on top of the analyzer to
perform the valve replacement.
5. Carefully, disconnect the 1/16 inch fitting on the pressure transducer.
6. Carefully, disconnect both 1/16 inch fittings on the moisture generator.
7. Remove the nuts holding the mounting bracket for the moisture gen-
erator. The moisture generator bracket, containing both the generator
and the pressure transducer, can now be lifted from the sample mani-
fold plate (Figure 8-5).
Figure 8-5.
Model 3050-RM moisture
generator.
NOTE
Содержание 3050-RM
Страница 11: ... xi ...
Страница 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Страница 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Страница 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Страница 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Страница 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Страница 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Страница 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...