8-12 | 3050 RM Moisture Analyzer
Maintenance of the Electronic Module
The electronics module was designed to easily identify and service the
boards and power supplies used in the Model 3050 RM Analyzer. These
components include:
• MCU Assembly
• Analog Interface Assembly
• Power Supply
• Fan
Removing the Electronic Module from the 3050 RM
Disconnect all power to the analyzer before removing the electronic
module.
Tools required
• 1/4” nut driver
• 1/16” hex wrench
• Phillips and slotted screwdrivers
Procedure for removing the electronics module
Use Figure 8-9 as a reference for locating the items described in this proce-
dure.
• Locate the four (4) Phillips head screws on the electronics module.
• Using the appropriate tool, remove the mounting hardware.
• With all the interconnecting wiring and tubing still attached, remove
the electronic module from the 3050 RM assembly by pulling up and
slightly towards the front of the analyzer.
• Once you remove the electronics module from the assembly, you can
place it on the analyzer cabinet as a resting fixture. However, take
care not to scratch the analyzer cabinet with the electronics module
(a piece of cardboard can be placed under the electronics module to
protect the cabinet).
All of the components are now accessible for maintenance or replacement.
!
CAUTION
Содержание 3050-RM
Страница 11: ... xi ...
Страница 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Страница 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Страница 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Страница 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Страница 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Страница 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Страница 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...