3-8 | 3050-RM Moisture Analyzer
Installing the Dryer (Figures 3-4a and b)
Prior to operating the 3050-RM analyzer, the dryer must be installed. A
dryer is supplied with this analyzer. Make sure that you have the new
dryer and the new VCR gaskets available before beginning installation.
The amount of time that the dryer fitting is left open should be kept to
less than a minute to avoid damage to the device.
1. On the right side of the rear of the analyzer is the dryer acess com-
partment (Figure 3-4).
2. Loosen the four captured screws on the door of the compartment to
remove the door.
3. If you have a single dryer analyzer, you must install the dryer in the
bottom position.
An optional second dryer or contaminant trap is available. If you
have the two-dryer or dryer/trap version of the 3050-RM, follow the
instructions for the one-dryer operation and then repeat for the second
dryer or trap.
!
CAUTION
Figure 3-4a.
Dryer locations for installation.
NOTE
Содержание 3050-RM
Страница 11: ... xi ...
Страница 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Страница 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Страница 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Страница 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Страница 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Страница 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Страница 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...