Service & Parts
|
8-11
!
CAUTION
Installing the new dryer / trap
Minimize the time that the dryer is exposed to the atmosphere after
removing the protective caps. AMETEK recommends that the dryer
be exposed for less than a minute.
1. Insert the replacement dryer into the dryer port on the analyzer. Use
the new gaskets provided with the dryer in each of the fittings.
2. Tighten the fittings until they are finger tight.
3 Install the dryer mounting screws to secure the dryer to the bracket.
4. Tighten both of the fittings so that the nuts are 1/8
th
of a turn past fin-
ger tight.
5. If you have a second dryer or trap. Repeat Steps 1 through 4 for the
second dryer or trap.
6. Place the dryer compartment cover onto the rear of the analyzer and
secure with the four captured screws.
7. Enter the production code for the new dryer.
The dryer production code is entered only once even if your analyzer
is equipped with two dryers.
8. Power up the analyzer. Wait until the boot-up sequence is complete.
- Press
TEST/CONFIG.
Scroll down to “Production Codes” and press Enter.
- Select “Reference Dryer” and press Enter to enter the production
code. Press Enter.
- Select “Dryer Config” and choose either Single (for one dryer ana-
lyzer) or Double, if your analyzer is equipped with two dryers.
- Press
Cancel several times to reverse through the menu and re-
turn to the normal operating screen.
!
CAUTION
Содержание 3050-RM
Страница 11: ... xi ...
Страница 20: ...1 8 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 24: ...2 4 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 27: ...Installation Start Up 3 3 Figures 3 1c Model 3050 RM dimensions top and sides ...
Страница 44: ...4 2 3050 RM Moisture Analyzer Figure 4 2 Menu map for 3050 RM Moisture Analyzer ...
Страница 63: ...Serial Communications 5 7 Figure 5 6 Defined Failure Codes ...
Страница 64: ...5 8 3050 RM Moisture Analyzer Figure 5 7a Variable Table ...
Страница 65: ...Serial Communications 5 9 Figure 5 7b Variable Table ...
Страница 66: ...5 10 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 78: ...7 2 3050 RM Moisture Analyzer This page intentionally left blank ...
Страница 91: ...Service Parts 8 13 Figure 8 9 Electronic Module ...
Страница 98: ...PPmV Centigrade Fahrenheit lbs mnscf mg Nm3 ppmW ...