Installation Guide
Thermal Imaging System
NIR-B Borescope
Electrical Auto Retract
2.1 NIR-B Borescope Thermal Imager
Details of the camera can be found in the NIR-B Borescope Installation Guide.
2.2 Imaging Workstation Computer
Details of the computer can be found in the NIR-B Borescope Installation Guide.
2.3 ECU
Power
:
Vac
90-132/180-264 VAC, auto select
Iac
2.20/0.83 A 115/230 VAC typical
Freq 50-60 Hz
Temperature :
0 °C to 50 °C
Size
:
800 x 500 x 300 mm nominal max
Weight
:
30 kg max
User I/O
:
Lens temperature
4-20mA, 0-200 ºC
Max loop Resistance 300 ohms
Isolated
Lens ‘over-temperature warning’ relay
Operating point: 150 ºC lens temperature
[retract operates at 180 ºC]
Contacts:
NO
Rating:
0.3 A at 125 VAC; 1 A at 30 VDC; resistive
‘System OK’ relay
‘In position’ relay
Contacts:
NO/NC
Rating:
10 A at 250 VAC; 10 A at 30 VDC; resistive
Sealing
:
to the requirements of IP54
Battery
:
Siemens 6EP19356MD31 [24 V, 2.5 Ah]
High Temperature Maintenance Free Sealed Lead Acid
2 - 0
SYSTEM COMPONENT
SPECIFICATIONS