Installation Guide
9 - 6
Thermal Imaging System
NIR-B Borescope
Electrical Auto Retract
Fig. 9-11 WaterSen Parameter
8)
Press the
ESC
key to navigate back to the screen of Fig 9.7. Then press the UP
key to return to normal operation.
9.4 LED Status (ECU)
Colour Function
On
Off
BLUE
MAINS
ECU has power from the mains
supply.
ECU power interrupted. Note:
the ECU & AR system will
continue to operate on battery
for a further 60 seconds.
GREEN
IN POSITION
AR slide is in its forward working
position.
The AR slide is moving, or in its
home position.
RED
FAULT
Indicates a fault condition has
occurred, or is preventing the
normal operation of of the slide.
Consult the Diagnostics section
Normal operation.
YELLOW BATTERY-CHARGE Internal battery is not at full
charge. The unit may not
operate correctly in the event of
a mains operation failure.
Internal battery fully charged
and ready.
6)
Using the
UP / DOWN
arrows, select
WaterSen
and press
OK
.
Fig. 9-10 WaterSen option
7)
Using the UP / DOWN arrows, adjust the T parameter to the desired time.
Fig. 9-11 shows the nominal 30 seconds.
NEVER set to more than 3 minutes
.
Press
OK
.