15.
Remove the FX3G-EEPROM-32L module from the expansion board by gently pulling on the
front cover.
16.
Install the plastic cover back on to the expansion board.
17.
Place the PLC RUN/STOP switch in the RUN position.
18.
Turn the panel power ON.
19.
Verify that the PLC Power Status LED is ON.
20.
Verify the PLC Run Status LED is ON.
21.
Shut and secure the panel front door.
22.
Navigate to the unit’s SOFTWARE VERSIONS HMI screen. Verify that the PLC software version
matches the version upgrade documentation.
NAMES AND FUNCTIONS OF PARTS:
PLC
[1] PLC Power Status LED:
Illuminated green while power is on the PLC.
[2] PLC Run Status LED:
Illuminated green while the PLC is running.
[3] RUN/STOP switch:
Starts or stops PLC code execution.
[4] Optional Equipment Connector: FX3G-EEPROM-32L module connection point.
FX3G-EEPROM-32L module
[3]
[4]
[1]
[2]