20
98-0156 Rev G
Press
“DAILY DEFAULTS”
.
For every day of the week, Monday-
Sunday, the disinfection program times
and conditions for the distribution
loops and the dialysis machines will
need to be entered. For example press
“MONDAY”
. The following screen will
be displayed; identical screens exist for
the other days of the week.
Firstly, enter the “
NUMBER OF
MACHINES
”. This indicates the number
of dialysis machines that will take
water for disinfection on this day. Next,
enter the
“HEATSAN START”
time. To
change the time, select each digit to be
changed individually. A pop up number
keypad will appear on the screen,
select the correct number and press
“ENTER”
.
After entering the Heatsan start time
enter a time that the water system
must be available for normal service.
Press the desired digit next to
“SERVICE”
and input correct values
until the correct time is displayed.
NB:
If NO heat disinfection is to take
place on a particular day enter 00:00
for the
“HEATSAN START”
time.
Содержание Heatsan 00HS208
Страница 51: ...50 98 0156 Rev G Heatsan Process Flow Diagram ...
Страница 52: ...51 98 0156 Rev G Wiring Diagrams ...
Страница 53: ...52 98 0156 Rev G ...
Страница 54: ...53 98 0156 Rev G ...
Страница 55: ...54 98 0156 Rev G ...
Страница 56: ...55 98 0156 Rev G ...
Страница 57: ...56 98 0156 Rev G ...
Страница 58: ...57 98 0156 Rev G Installation Layout Water Connection Detail ...
Страница 62: ...61 98 0156 Rev G Tubing with Heatsan and MediQA Hidden ...
Страница 64: ...63 98 0156 Rev G RO Divert Valve and Heatsan Divert Valve 1 Heatsan Divert Valve 2 ...
Страница 65: ...64 98 0156 Rev G Heatsan Discharge Solenoid SV3A NC Heatsan Fill Solenoid SV3B NO ...
Страница 70: ...69 98 0156 Rev G ...