Refer to Section 6.10 for FACTORY
SETUP PAGE 2 values.
DRAW OFF,
is the volume of water, in
gallons, each dialysis machine requires
for disinfection.
RATE
in US gal/min that each machine
draws water from the loop.
Press
“NEXT PAGE”
to enter
FACTORY
SETUP PAGE 3
.
Refer to Section 6.10 for FACTORY
SETUP PAGE 3 values.
TIME PER DIALYSIS MACHINE,
in
hrs/min. This is the time each machine
takes to carry out a heat disinfection
cycle.
Press
“NEXT PAGE”
to enter
FACTORY
SETUP PAGE 4
.
Refer to Section 6.10 for FACTORY
SETUP PAGE 4 values.
COOL DOWN DROP
, is the pre-set
value that controls the opening of the
cold water fill valve during cool down
cycle.
Press
“NEXT PAGE”
to enter
FACTORY
SETUP PAGE 5
.
Refer to Section 6.10 for FACTORY
SETUP PAGE 5 values.
WATER FEED ON,
is the temperature
at which the fill water feed is turned
on.
WATER FEED OFF
, during “dual”
disinfection refill stage this pre-set
value controls the filling of the tank,
Содержание Heatsan 00HS208
Страница 51: ...50 98 0156 Rev G Heatsan Process Flow Diagram ...
Страница 52: ...51 98 0156 Rev G Wiring Diagrams ...
Страница 53: ...52 98 0156 Rev G ...
Страница 54: ...53 98 0156 Rev G ...
Страница 55: ...54 98 0156 Rev G ...
Страница 56: ...55 98 0156 Rev G ...
Страница 57: ...56 98 0156 Rev G ...
Страница 58: ...57 98 0156 Rev G Installation Layout Water Connection Detail ...
Страница 62: ...61 98 0156 Rev G Tubing with Heatsan and MediQA Hidden ...
Страница 64: ...63 98 0156 Rev G RO Divert Valve and Heatsan Divert Valve 1 Heatsan Divert Valve 2 ...
Страница 65: ...64 98 0156 Rev G Heatsan Discharge Solenoid SV3A NC Heatsan Fill Solenoid SV3B NO ...
Страница 70: ...69 98 0156 Rev G ...