62
098-0007 Rev B
Tank Full
FS1: Closed
FS2: Open
1
FS1 refers to high level float switch
2
FS2 refers to low level float switch
12.5
Setting Current Switch
The Centurion Plus is equipped with a current switch that monitors the current draw from the device to
prevent the heater from activating when the current draw exceeds 11 A. This is intended to prevent the
MROC and Centurion Plus from operating their heating element at the same time.
NOTE:
This feature only works if the MROC is plugged into the power outlet located at the rear of the
Centurion Plus Heat disinfection device.
CAUTION:
The system must be powered on when adjusting the current switch. Only trained personnel
should attempt this adjustment.
Steps
Actions
Step 1:
Ensure the feed water RO is powered on
and operating. Navigate to diagnostics
menu page 1 and manually activate the
tank fill solenoid valve (SV1).
Step 2:
Allow the tank to fill for approximately 2
minutes. Verify that the bottom float
switch (FS2) is shown as “Open” on
diagnostics page 2. Manually shut off
valve SV1.
Step 3:
Remove the side covers from the device
and unscrew the fasteners for the top
panel shown. Carefully move the top
panel to the side to expose the current
switch.
Содержание 000-HS120
Страница 2: ......
Страница 24: ...22 098 0007 Rev B Right Side...
Страница 25: ...23 098 0007 Rev B Left Side Top with top panel and controller removed...
Страница 27: ...25 098 0007 Rev B 6 3 Flow Schematic 6 3 1 Standby Mode...
Страница 28: ...26 098 0007 Rev B 6 3 2 Preheat Mode...
Страница 29: ...27 098 0007 Rev B 6 3 3 Dispensing Mode...
Страница 30: ...28 098 0007 Rev B 6 3 4 Drain Mode...
Страница 68: ...66 098 0007 Rev B 13 SPARE PART LISTING 13 1 Exploded View...
Страница 69: ...67 098 0007 Rev B...
Страница 70: ...68 098 0007 Rev B...
Страница 71: ...69 098 0007 Rev B...
Страница 72: ...70 098 0007 Rev B...
Страница 73: ...71 098 0007 Rev B...
Страница 74: ...72 098 0007 Rev B...