33
098-0007 Rev B
7.5.2
System Operational Controls
The Centurion Plus Heat Disinfection device is controlled by a Programmable Logic Controller (PLC) with
an integrated operation panel that sends and receives signals to and from various components in the
device. The user can set and view various parameters on the Human Machine Interface (HMI) display on
the front of the device. This section provides an overview of each screen display to allow the user to gain
familiarity with the device and basic system parameters.
Screen Example
Screen Description
Normal Operation:
Storage Mode Active – no schedule:
Storage Mode Active – with set schedule:
Home Screen
From this screen the user can start a disinfection
cycle, enter the settings menu, see the last
disinfection status, or navigate to the alarm menu
if an alarm condition is active. The last
disinfection status will show the user the time and
date when the last disinfection was completed or
failed due to an Abort or Alarm condition. If
storage mode is activated, a message will be
displayed informing the user that the system will
operate based on a schedule.
This screen also displays the current time, date
and operating mode for the device. The current
time and date may be adjusted following the
Start
Pressing this button will prompt the user to
confirm that they would like to proceed with the
disinfection.
NOTE:
the start button will be hidden if storage
mode has been activated and at least one
day has been set for a scheduled
disinfection.
Schedule
Pressing this button will prompt the user to enter
a “Level 1” password to access the disinfection
schedule or system settings for the device.
Disinfection Status Message
The disinfection status message will display the
status of the last disinfection cycle. Pressing on
this message will allow the user to view the
disinfection log.
Storage Mode Active Message
The message will be displayed when the storage
mode is activated on the device. This message is
intended to inform the user if the device has a set
schedule.
Abort Scheduled Operation
Содержание 000-HS120
Страница 2: ......
Страница 24: ...22 098 0007 Rev B Right Side...
Страница 25: ...23 098 0007 Rev B Left Side Top with top panel and controller removed...
Страница 27: ...25 098 0007 Rev B 6 3 Flow Schematic 6 3 1 Standby Mode...
Страница 28: ...26 098 0007 Rev B 6 3 2 Preheat Mode...
Страница 29: ...27 098 0007 Rev B 6 3 3 Dispensing Mode...
Страница 30: ...28 098 0007 Rev B 6 3 4 Drain Mode...
Страница 68: ...66 098 0007 Rev B 13 SPARE PART LISTING 13 1 Exploded View...
Страница 69: ...67 098 0007 Rev B...
Страница 70: ...68 098 0007 Rev B...
Страница 71: ...69 098 0007 Rev B...
Страница 72: ...70 098 0007 Rev B...
Страница 73: ...71 098 0007 Rev B...
Страница 74: ...72 098 0007 Rev B...