17
098-0007 Rev B
4.9
Performing first disinfection cycle
Prior to commissioning the Centurion Plus Heat disinfection device for dialysis use, an initial disinfection
cycle shall be preformed to ensure the system functions as intended. All Centurion Plus Heat disinfection
devices are functionally tested by AmeriWater prior to shipment. Refer to Section 8.3 for detailed
information on performing a disinfection using the Centurion Plus Heat disinfection device.
4.10
Pre Dialysis Checks
Prior to using the device for dialysis treatment, it is recommended that the following checks are
preformed to verify the device will function as designed:
Check that the electrical mains lead is securely inserted in place on the unit and at the wall
socket. Verify that the lead is not damaged or likely to cause anyone to trip over it.
All the water connections are in place, not kinked and show no signs of leaks.
The “Blue” water tubing is connected to the “In/Entreé” port (Centurion), the “Black” water
tubing is connected to the “System Drain” port (Centurion Plus Heat disinfection device).
The water supply is turned on to the Centurion.
There are no warning messages displayed on the touch-screen. (Refer to Section 11, if any
alarm or warning messages are displayed.
Water should pass through the Centurion Plus Heat disinfection device as if it were not installed
while the Centurion is operating and the Centurion Plus Heat disinfection device is not in a
disinfection cycle.
Check that the Centurion is operating properly before attempting to use the RO with the
Centurion Plus Heat Disinfection device.
4.11
Decommissioning for Relocation
Prior to moving the Centurion Plus Heat disinfection device, the following steps should be followed to
ensure the device is safely transported.
Ensure the device is in “Standby Mode” and on the Home Screen.
RO reject
water outlet
RO product
water return
RO product
water outlet
Feed water inlet (From
pretreatment outlet)
RO water inlet
(From RO)
RO water
return (To RO)
RO water outlet to
HD machine (Product
Outlet)
RO water return from
HD machine (Product
Return)
Common Drain Outlet
(System Drain)
RO drain inlet
(RO Drain)
Содержание 000-HS120
Страница 2: ......
Страница 24: ...22 098 0007 Rev B Right Side...
Страница 25: ...23 098 0007 Rev B Left Side Top with top panel and controller removed...
Страница 27: ...25 098 0007 Rev B 6 3 Flow Schematic 6 3 1 Standby Mode...
Страница 28: ...26 098 0007 Rev B 6 3 2 Preheat Mode...
Страница 29: ...27 098 0007 Rev B 6 3 3 Dispensing Mode...
Страница 30: ...28 098 0007 Rev B 6 3 4 Drain Mode...
Страница 68: ...66 098 0007 Rev B 13 SPARE PART LISTING 13 1 Exploded View...
Страница 69: ...67 098 0007 Rev B...
Страница 70: ...68 098 0007 Rev B...
Страница 71: ...69 098 0007 Rev B...
Страница 72: ...70 098 0007 Rev B...
Страница 73: ...71 098 0007 Rev B...
Страница 74: ...72 098 0007 Rev B...