11
098-0007 Rev B
4.3
Installation on Centurion Pretreatment cart
The Centurion Plus Heat Disinfection Device is designed to be installed onto a cart with the AmeriWater
MROC Heat Disinfection RO system. The Cart contains the sediment and carbon filtration required for the
RO system to perform as intended. The following steps are required to install the Centurion Plus Heat
disinfection device onto the Cart along with the MROC.
a)
Remove the right hand side cover on the MROC and check the P17 and P18 check valves that are
installed in the device. If the valves are stamped ACV-4FFSS-V-15 and ACV-4FFSS-V-20, proceed
to the next step, if the valves are stamped ACV-4FFSS-V-25 and ACV-4FFSS-V-35, proceed to
step d.
b)
Disconnect the tube connections shown and remove the existing check valves. Water will leak
from the Y connector when disconnected. Be sure to drain water out of lower pan of Centurion
after removing the valves.
c)
Install the check valves included with the install kit. The valve labeled ACV-4FFSS-V-35 must be
installed on the permeate return to tank line as shown. Failure to do this will cause the RO
product water to be diverted to the break tank on the RO.
Содержание 000-HS120
Страница 2: ......
Страница 24: ...22 098 0007 Rev B Right Side...
Страница 25: ...23 098 0007 Rev B Left Side Top with top panel and controller removed...
Страница 27: ...25 098 0007 Rev B 6 3 Flow Schematic 6 3 1 Standby Mode...
Страница 28: ...26 098 0007 Rev B 6 3 2 Preheat Mode...
Страница 29: ...27 098 0007 Rev B 6 3 3 Dispensing Mode...
Страница 30: ...28 098 0007 Rev B 6 3 4 Drain Mode...
Страница 68: ...66 098 0007 Rev B 13 SPARE PART LISTING 13 1 Exploded View...
Страница 69: ...67 098 0007 Rev B...
Страница 70: ...68 098 0007 Rev B...
Страница 71: ...69 098 0007 Rev B...
Страница 72: ...70 098 0007 Rev B...
Страница 73: ...71 098 0007 Rev B...
Страница 74: ...72 098 0007 Rev B...