60
098-0007 Rev B
HD machine low water
alarm when Centurion Plus
is in standby mode
Valve SV4 fails closed
Verify that valve SV4 is not
manually activated on
diagnostics menu page 1
Verify that RO is producing
approximately 1500 mL/min
using the sample port on the
RO product hose
Valve SV1 fails open to tank
Use diagnostics menu to
check tank level
Verify that valve SV1 is not
manually activated in
diagnostics menu
Verify that water is not
exiting the Centurion Plus via
the over flow port
Valve SV3 fails open to drain
Disconnect RO reject line
from Centurion Plus. Verify
water does not pass through
system drain port
Rear fan does not function
(Fan runs at all times when
the rocker switch in on
position)
12 VDC power supply (PS2)
failure
Locate power supply labeled
PS2 in control panel
Check for 12 VDC between
pins 2 and 4 on power
supply
Mechanical failure of fan
Verify 12 VDC power supply
functionality
Check that connector is
plugged in
Check that fan wires are not
loose
Leak at push fit connections
Tube end not cut square
Remove tube from
connector. Verify end of tube
is cut square
Tube not fully inserted
Verify that tube or fitting is
fully inserted into fitting
Содержание 000-HS120
Страница 2: ......
Страница 24: ...22 098 0007 Rev B Right Side...
Страница 25: ...23 098 0007 Rev B Left Side Top with top panel and controller removed...
Страница 27: ...25 098 0007 Rev B 6 3 Flow Schematic 6 3 1 Standby Mode...
Страница 28: ...26 098 0007 Rev B 6 3 2 Preheat Mode...
Страница 29: ...27 098 0007 Rev B 6 3 3 Dispensing Mode...
Страница 30: ...28 098 0007 Rev B 6 3 4 Drain Mode...
Страница 68: ...66 098 0007 Rev B 13 SPARE PART LISTING 13 1 Exploded View...
Страница 69: ...67 098 0007 Rev B...
Страница 70: ...68 098 0007 Rev B...
Страница 71: ...69 098 0007 Rev B...
Страница 72: ...70 098 0007 Rev B...
Страница 73: ...71 098 0007 Rev B...
Страница 74: ...72 098 0007 Rev B...