HEAT DISINFECTION
8.1
What is being disinfected?
The Centurion Plus Heat disinfection device is intended to disinfect the hose used to feed a Hemodialysis
machine with purified RO water. Typically, the RO and Hemodialysis machine are disinfected using their
own automated disinfection processes. Neither the RO or Hemodialysis machines disinfection procedures
disinfect the feed water hose which leaves it up to the end user to chemically disinfect. The Purpose of
the Centurion Plus Heat disinfection device is to disinfect this hose between the RO and Hemodialysis
machine using an automated heat disinfection cycle.
8.2
Frequency
Disinfection of the feed water hose to the Hemodialysis machine is required monthly. AmeriWater
recommends at least one weekly heat disinfection of the feed water hose for the hemodialysis machine.
Environmental conditions may warrant more frequent disinfections.
8.3
Heat Disinfection Sequence
Screen Display
Operators Action
Step 1:
From the Home Screen select “Start”
Step 2:
Pressing “Start” will cause the confirmation
screen to appear. Select “Yes” to begin the
disinfection cycle. At this point, the RO system
feeding the Centurion Plus Heat disinfection
device should be operating.
Step 3:
The Preheat cycle will begin upon confirming the
disinfection. At the beginning of this cycle, the
storage tank will begin to fill with RO water. Once
the tank is full, the RO water will continue to pass
through the device as if it were in Standby.
The preheat cycle will remain active until the
water in the tank reaches the desired disinfection
temperature. At any time during the preheat
cycle, the Hemodialysis machine can be placed
into a rinse cycle (rinse cycle must allow for at
Содержание 000-HS120
Страница 2: ......
Страница 24: ...22 098 0007 Rev B Right Side...
Страница 25: ...23 098 0007 Rev B Left Side Top with top panel and controller removed...
Страница 27: ...25 098 0007 Rev B 6 3 Flow Schematic 6 3 1 Standby Mode...
Страница 28: ...26 098 0007 Rev B 6 3 2 Preheat Mode...
Страница 29: ...27 098 0007 Rev B 6 3 3 Dispensing Mode...
Страница 30: ...28 098 0007 Rev B 6 3 4 Drain Mode...
Страница 68: ...66 098 0007 Rev B 13 SPARE PART LISTING 13 1 Exploded View...
Страница 69: ...67 098 0007 Rev B...
Страница 70: ...68 098 0007 Rev B...
Страница 71: ...69 098 0007 Rev B...
Страница 72: ...70 098 0007 Rev B...
Страница 73: ...71 098 0007 Rev B...
Страница 74: ...72 098 0007 Rev B...