![AmeriWater 000-HS120 Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/ameriwater/000-hs120/000-hs120_operation-and-maintenance-manual_3043934036.webp)
34
098-0007 Rev B
Screen Example
Screen Description
This button will be hidden unless the device is in
storage mode AND the current day is equal to the
scheduled day of disinfection. Pressing and
holding the button (2 seconds) will cancel a
scheduled disinfection until the following day.
Disinfection Confirmation
This screen will appear after the “START” button
is pressed on the home screen. Pressing the “NO”
button will return to the Home Screen. Pressing
the “Yes” button will begin a disinfection cycle.
A warning message appears at the bottom of the
screen which informs the user that the
disinfection should not be started while dialysis
treatment is occurring. The dialysis machine
should be connected and in a “Rinse” mode to
accept the hot water once the preheat cycle ends.
Temperature Maintain Mode Disabled:
Temperature Maintain Mode Active:
Preheat Mode Operation Screen
This screen appears once the user presses “Yes”
on the confirmation screen. On this screen, the
current tank temperature is displayed. A progress
bar will scroll from left to right as the temperature
increases. When the disinfection temperature is
reached, the progress bar will be at 100%.
A warning message will appear on the screen
above the abort button when the disinfection
temperature is set below 176 °F.
If the “Preheat Delay Timer” is active, a
message will appear on the screen with a
calculated draw off time. This informs the user of
the time when the HD machine should be ready
to draw in hot water.
Abort
The abort button will send the user to an abort
confirmation screen. When the abort is confirmed,
the user will be sent back to the Home Screen.
The abort will be logged in the Alarm History Log.
Settings
The settings menu at the bottom left corner of
the screen will allow the user to access the
settings menu via a “Level 1” password. This is
intended for use by technicians when diagnosing
issues with the device.
Hot Water to HD
Содержание 000-HS120
Страница 2: ......
Страница 24: ...22 098 0007 Rev B Right Side...
Страница 25: ...23 098 0007 Rev B Left Side Top with top panel and controller removed...
Страница 27: ...25 098 0007 Rev B 6 3 Flow Schematic 6 3 1 Standby Mode...
Страница 28: ...26 098 0007 Rev B 6 3 2 Preheat Mode...
Страница 29: ...27 098 0007 Rev B 6 3 3 Dispensing Mode...
Страница 30: ...28 098 0007 Rev B 6 3 4 Drain Mode...
Страница 68: ...66 098 0007 Rev B 13 SPARE PART LISTING 13 1 Exploded View...
Страница 69: ...67 098 0007 Rev B...
Страница 70: ...68 098 0007 Rev B...
Страница 71: ...69 098 0007 Rev B...
Страница 72: ...70 098 0007 Rev B...
Страница 73: ...71 098 0007 Rev B...
Страница 74: ...72 098 0007 Rev B...