CAL2 GAIN ERR
- Calibration point 2 gain error. See
CAL GAIN ERR
.
•
CAL2 RESt ERR
- Calibration point 2 resistance error. Indicates that the calibration
resistance constant has been calculated at either a negative value or a value greater than 20Ω
(i.e. excessive lead resistance). Check cell connections and
CAL2
calibration procedure.
Note:
It is essential in conductivity measurement that the resistance across the cell is always
greater than 80Ω. If the resistance is less than this then it may be necessary to use a cell with a
higher cell constant e.g. it may be necessary to change from K=0.1 to K=1. The resistance at any
given conductivity level can be found from the formula:
Resistance
(
Ohms
) =
1
Conductivity /cm
×
K f actor
e.g. for a K=0.1 cell in a 2000 uS/cm solution the resistance is 50 Ohms (see below) which is not
acceptable. Changing to a K=1 cell would result in an acceptable resistance of 500Ω.
Resistance
=
1
2000
×
10
−
6
×
0
.
1 = 50
Ohms