mS.m
to display in milliSiemens per metre (mS/m)
Note that if the conductivity display units are changed the instrument must be recalibrated.
5.38
Conductivity uncalibration
Display:
UCAL cond
Range:
n/a
Default Value:
n/a
This function is used to set the conductivity calibration back to the factory calibration values and
will be seen only if the
SEt dISP
function is set to
cond
. This function should only be used
when calibration problems exist and it is necessary to clear the calibration memory. See section
for details.
5.39
Resistivity uncalibration
Display:
UCAL RESt
Range:
n/a
Default Value:
n/a
This function is used to set the resistivity calibration back to the factory calibration values and
will be seen only if the
SEt dISP
function is set to
RESt
. This function should only be used
when calibration problems exist and it is necessary to clear the calibration memory. See section
for details.
5.40
ppm uncalibration
Display:
UCAL ppm
Range:
n/a
Default Value:
n/a
This function is used to set the ppm calibration back to the factory calibration values and will
be seen only if the
SEt dISP
function is set to
ppm
. This function should only be used when
calibration problems exist and it is necessary to clear the calibration memory. See section
for
details.
5.41
Percentage uncalibration
Display:
UCAL Pcnt
Range:
n/a
Default Value:
n/a
This function is used to set the resistivity calibration back to the factory calibration values and
will be seen only if the
SEt dISP
function is set to
Pcnt
. This function should only be used
when calibration problems exist and it is necessary to clear the calibration memory. See section
for details.