may be set to operate from either the conductivity, resistivity, ppm, sodium chloride % by weight,
temperature values. Note only the mode selected at the
SEt OPER
function or the temperature
can be selected. Select
cond
for conductivity,
RESt
for resistivity,
ppm
for
ppm
, or
Pcnt
for sodium chloride % or
"C
for temperature. Set the
bAr_
and
bAr~
functions to suit the
output mode selected. Note: if set for conductivity, resistivity or ppm then the bargraph will go
to full scale if the display goes to overrange (
-or-
displayed on the 7 segment display). If set for
temperature an overrange display will cause the bargraph to revert to the default temperature set
at the
dEF "C
function i.e. if required the default temperature can be set to a value which will
give required bargraph display if an overrange, e.g. open circuit, occurs.
5.51
Analog output mode
Display:
REC
Range:
cond, RESt, PPM, Pcnt
or
"C
Default Value:
cond
Analog output mode - seen only when the analog output option is fitted. The analog output
may be set to operate from either the conductivity, resistivity, ppm, sodium chloride % by weight,
temperature values. Note only the mode selected at the
SEt OPER
function or the temperature
can be selected. Select
cond
for conductivity,
RESt
for resistivity,
ppm
for
ppm
, or
Pcnt
for
sodium chloride % or
"C
for temperature. Set the
REC_
and
REC~
functions to suit the output
mode selected. Note: if set for conductivity, resistivity or ppm then the output will go to full scale
if the display goes to overrange (
-or-
displayed on the 7 segment display). If set for temperature
an overrange display will cause the output to revert to that for the default temperature set at the
dEF "C
function i.e. if required the default temperature can be set to a value which will give
required analog output if an overrange, e.g. open circuit, occurs.
5.52
Second analog output mode
Display:
REC2
Range:
cond, RESt, PPM, Pcnt
or
"C
Default Value:
cond
Analog output mode - seen only when the dual analog output option is fitted. The second analog
output may be set to operate from either the conductivity, resistivity, ppm, sodium chloride %
by weight, temperature values. Note only the mode selected at the
SEt OPER
function or the
temperature can be selected. Select
cond
for conductivity,
RESt
for resistivity,
ppm
for
ppm
, or
Pcnt
for sodium chloride % or
"C
for temperature. Set the
REC_
and
REC~
functions
to suit the output mode selected. Note: if set for conductivity, resistivity or ppm then the output
will go to full scale if the display goes to overrange (
-or-
displayed on the 7 segment display).
If set for temperature an overrange display will cause the output to revert to that for the default
temperature set at the
dEF "C
function i.e. if required the default temperature can be set to a
value which will give required analog output if an overrange, e.g. open circuit, occurs.