Functions in this second table are available only in
CAL
mode or if
ACCS
is set to
ALL
Display
Function
Range
Default
Your
record
Ref/Page
bAr
tYPE
Bargraph type for instruments
with bargraph display
bAr
,
S.dot
,
d.dot
,
C.bAR
or
r.dot
bAr
/
dGOP
Digital output option mode
(
∗
Optional
)
bcd
,
b.SCL
,
bin
or
bin2
bin2
/
dG.OP
Digital output option polarity
(
∗
Optional
)
AIo
or
AHi
AIo
/
bcd
Strt
Digital output option BCD start
position (
∗
Optional
)
0
,
1
or
2
0
/
diG_
Digital output option low value
(
∗
Optional
)
Any display
value
0
/
diG~
Digital output option high value
(
∗
Optional
)
Any display
value
1000
/
drnd
Display rounding
1
to
5000
1
/
dCPt
Decimal point
0
,
0.1
etc.
0
/
FLtr
Digital filter
0
to
8
2
/
dOSE
PERd
Special dosing mode operation
0
to
600
0
/
"C
tYPE
Temperature sensor type
NONE, 100,
1000, L335,
t100
or
25J1
NONE
/
dEF
"C
Default temperature value
0.0
to
200.0
25.0
/
SOL
SLPE
Solution temperature
compensation slope
-6.00
to
0.00
-2.00
/
SOL
"C
Solution temperature
compensation reference
0.0
to
100.0
25.0
/
PRbE
CNSt
Cell K factor
0.01, 0.05,
0.1, 0.5, 1.0,
2.0, 5.0, 10,
50
or
100
0.1
/
H.OFF
or
H.ON
Hydrogen ion compensation
H.OFF
or
H.ON
H.OFF
/
Ind
PRbE
Inductive cell selection
on
or
OFF
OFF
/
CAL
NULL
Display null calibration
n/a
n/a
/
CAL1
First live input calibration
scaling point
Any display
value
n/a
/
(
∗
Optional
)—this function will only be accessible if the relevant option is fitted
PM4COMAN-2.2-1
15 of