5.42
Temperature uncalibration
Display:
UCAL "C
Range:
n/a
Default Value:
n/a
This function is used to set the temperature calibration back to the factory calibration values.
This function should only be used when calibration problems exist and it is necessary to clear the
calibration memory. See section
for details.
5.43
ppm conversion factor
Display:
PPm FACt
Range:
0.200
to
2.000
Default Value:
0.560
This function is used to display and sets the ppm conversion factor and will be seen only if the
SEt dISP
function is set to
ppm
. See section
for details.
5.44
Set display type
Display:
SEt dISP
Range:
cond, RESt, PPm
or
Pcnt
Default Value:
cond
This function is used to set the default display to either conductivity (
cond
) in the display units
selected or resistivity (
RESt
) in MΩ or parts per million (
ppm
) or Sodium Chloride (NaCl)
percentage by weight (
Pcnt
). Select the default display required for measurement.
5.45
Analog output PI control
Display:
rec ctrI
Range:
on
or
OFF
Default Value:
OFF
Analog output mode - seen only when analog output option is fitted. This function allows selection
of
on
or
OFF
for PI control analog output. If set to
OFF
the analog output operates as a
retransmission output and uses the functions described in this chapter. If set to
on
the analog
output operates as a PI control output.
When this function is set to
on
the following associated functions will appear:
C.SEt
,
C.SPN
,
C_PG
,
C_Po
,
C.IG
,
CIL.H
,
CIL.L
and
REC SPAC
. These functions are not detailed in
this manual. Refer to the separate “PM4 Panel Meter Optional Output Addendum” booklet for
description of the analog PI control functions and wiring details.
PM4COMAN-2.2-1