![AMALGAMATED INSTRUMENT PM4-CO Скачать руководство пользователя страница 24](http://html.mh-extra.com/html/amalgamated-instrument/pm4-co/pm4-co_instruction-manual_2915937024.webp)
function using the
^
or
v
button.
5.10
Analog output option high value
Display:
REC~
Range:
Any display value
Default Value:
1000
Seen only when analog retransmission option fitted. Refer to the separate “PM4 Panel Meter
Optional Output Addendum” booklet supplied when this option is fitted for wiring details and link
settings. Displays and sets the analog retransmission (4–20mA, 0–1V or 0–10V, link selectable)
output high display value (20mA, 1V or 10V) in displayed engineering units. To set the analog
output high value go to the
REC~
function and use the
^
or
v
push buttons to set the required
value then press
F
to accept this selection.
Example:
If it is required to retransmit 20mA when the display indicates
50
then select
50
in
this function using the
^
or
v
button.
5.11
Second analog output option low value
Display:
REC_ Ch2
Range:
Any display value
Default Value:
0
See
REC_
function
for description of operation.
5.12
Second analog output option high value
Display:
REC~ Ch2
Range:
Any display value
Default Value:
1000
See
REC~
function
for description of operation.
5.13
Bargraph low value
Display:
bAr_
Range:
Any display value
Default Value:
0
Seen only in bargraph display instruments. Displays and sets the bar graph low value i.e. the value
on the 7 segment display at which the bargraph will start to rise. This may be independently set
anywhere within the display range of the instrument. Note: The
bAr_
and
bAr~
settings are
referenced from the 7 segment display readings, not the bargraph scale values. The bargraph scale
may scaled differently to the 7 segment display. For example the bargraph scale may be indicating
percentage fill of a tank whilst the 7 segment display is indicating actual process units. To set
bargraph low level go to the
bAr_
function and use the
^
or
v
push buttons to set the value
required then press
F
to accept this value.
24 of
PM4COMAN-2.2-1