0.1
(1 decimal place),
0.02
(2 decimal places),
0.003
(3 decimal places) and
0.0004
for display
with more than 4 digits. Note if the decimal point is altered the display will need to be recalibrated
and alarm etc. settings checked.
5.23
Digital filter
Display:
FLtr
Range:
0
to
8
Default Value:
2
Displays and sets the digital filter value.
Digital filtering uses a weighted average method of
determining the display value and is used for reducing display value variation due to short term
interference. The digital filter range is selectable from
0
to
8
, where
0
= none and
8
= most
filtering. Use
^
or
v
at the
FLtr
function to alter the filter level if required. Note that the
higher the filter setting the longer the display may take to reach its final value when the input is
changed, similarly the relay operation and any output options will be slowed down when the filter
setting is increased. To set the digital filter value go to the
FLtr
function and use the
^
or
v
push buttons to set the required value then press
F
to accept this selection.
5.24
Special “blowdown” dosing mode operation
Display:
dOSE PERd
Range:
0
to
600
Default Value:
0
Special dosing operation mode - sets relay 1 operation period for special “blowdown” dosing pump
operation mode. This function can be set from 0 to 600 seconds. This function will only be seen
if function
A1
is set to
dOSE
. Refer to the
A1,A2
etc. function for a description of operation of
this special mode.
5.25
Temperature sensor type
Display:
"C tYPE
Range:
NONE, 100, 1000, L335, t100
or
25J1
Default Value:
NONE
Temperature sensor type used, note see also electrical link settings in chapter
Choices available are:
•
NONE
- no temperature sensor used
•
100
- Pt100 temperature sensor used
•
1000
- Pt100 temperature sensor used
•
L335
- LM335 temperatures sensor used
•
t100
- 100Ω thermistor sensor used
•
25J1
- UUB25J1 thermistor sensor used
28 of
PM4COMAN-2.2-1